Operator’s Manual ACF-NT, ACF-NT V0309
42/23-572 EN Rev. 5
“Multi-FID” Screen
Figure 7
“Multi-FID” Screen
Indication
The “Multi-FID” screen displays the values of the VOC analyzer temperature and
pressure regulators.
Parameter
Default
T-Co.D
Detector temperature (on “ASP Module” screen)
180 °C
T-Co.E
Sample gas port temperature
180 °C
Flame
Flame temperature
260–300 °C
A.Pres
Combustion air pressure
725 hPa
H2
Combustion gas pressure
1100 hPa
Input
Instrument air pressure at combustion chamber
inlet
700 hPa
Output
Instrument air pressure at combustion chamber
outlet
600 hPa