62
18. COMMANDS
18-1 Command List
Commands to query weighing data
C
Cancels
the
S
or
SIR
command.
Q
Requests the weighing data immediately.
S
Requests the weighing data when stabilized.
SI
Requests the weighing data immediately.
SIR
Requests the weighing data continuously.
E
SC
P
Requests the weighing data when stabilized.
Note: The “Q” and “SI” commands, the “S” and “
E
SC
P” commands behave the same.
Commands to control the balance
CAL
Same as the CAL key.
MCL
Deletes all data in memory.
MD:nnn
Deletes weighing data with the data number nnn.
OFF
Turns the display off.
ON
Turns the display on.
P
Same as the ON:OFF key
PT:
∗∗∗
.
∗∗
g
Sets the tare value.
The unit added is the unit that is output in the A&D standard format. For
the counting or percent mode, gram is used. Set the tare value so that
the total value of the sample and the tare does not exceed the weighing
capacity. Negative values can not be used.
?PT
Requests the tare value.
Outputs the tare value set by the TR or PT: command.
R
Same as the RE-ZERO key
SMP
Same as the SAMPLE key.
U
Same as the MODE key
T
Same as the RE-ZERO key
TR
Tares the balance (Cancels the container’s weight.)
Available only when the load on the weighing pan is greater than the
zero point. The zero point is set by the R, T or Z command, by pressing
the RE-ZERO key, or using the external input.
Z
Same as the RE-ZERO key
E
SC
T
Same as the RE-ZERO key
Note: The “R”, “T”, “Z” and “
E
SC
T” commands behave the same.
Commands to control the memory function
?MA
Outputs all data in memory.
?MQnnn
Outputs weighing data with the data number nnn.
?MX
Outputs the number of weighing data in memory.
MCL
Deletes all data in memory.
MD:nnn
Deletes weighing data with the data number nnn.
UN:mm
Changes the unit mass stored in memory with the number of mm (01-20).
?UN
Outputs the unit mass number of the selected unit mass.
UW:
∗∗∗
.
∗∗
g
Sets the unit mass value. e.g.,UW:+0.123 g (to set the unit mass to
0.123 g: represents a space.)
?UW
Outputs the unit mass value of the selected unit mass number.
?ID
Requests the ID number.
?SN
Requests the serial number.
?TN
Requests the model name.