70
Display Error code
Description
Weighing pan Error
The weight value is too light.
Confirm that the weighing pan is properly installed and
calibrate the balance.
Sample mass error
The balance can not store the sample for the counting
mode or for the percent mode because it is too light.
Use a larger sample.
Unit mass error
The sample unit mass for the counting mode is too
light. Storing and using it for counting will cause a
counting error.
Add samples to reach the specified number and press
the PRINT key.
Pressing the PRINT key without adding samples will
shift the balance to the counting mode. But, for
accurate counting, be sure to add samples.
(CHECK NO)
Internal error
Indicates an internal error as the result of self-check
function.
Repair is required. Contact the local A&D dealer.
(Blinking)
Memory full
The amount of weighing data in memory has reached the
maximum capacity.
Delete data in memory to store new data. For details, refer
to “12. DATA MEMORY”.
Memory type error
Type of memory set in the function table and type of
data stored are different.
For details, refer to “12. DATA MEMORY”.
EC, E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC, E01
Undefined command error
An undefined command was received.
Confirm the command.
EC, E02
Not ready
A received command can not be processed.
e.g. The balance received a Q command, but not in the
weighing mode.
e.g. The balance received a Q command while
processing a RE-ZERO command.
Adjust the delay time to transmit a command.