67
Setting a negative target value and filling with a sample until the display becomes zero
18-4 Control Using CTS and RTS
Depending on the “
Ct5
”
parameter of “Serial interface (
5if
)”, the balance performs as follows:
Ct5 0
Regardless of whether the balance can receive a command or not, the balance keeps the CTS line
HI. The balance outputs data regardless of the condition of the RTS line.
Ct5 1
The CTS line is kept Hi normally. When the balance can not receive the next command (e.g. while
the balance is processing last command), the balance sets the CTS line to Lo. The balance confirms
the level of the RTS line before outputting a set of data. If the RTS level is Hi, the balance outputs
data. If the RTS level is Lo, data is not output (The data is canceled).
18-5 Settings Related to RS-232C
Concerning the RS-232C, the balance has two functions: “Data output (
dout
)” and “Serial interface
(
5if
)”. Set each function as necessary.
Place a container
Zero display
Personal computer Time Balance
R command
CR
R
LF
AKCR LF
AKCR LF
CR LF
g
0
0
0
0
1
:
T
P
.
PT: command to set a target weight of 100 g
Before command
execution
AKCR LF
AKCR LF
Target weight display
Fill with sample until the
display becomes zero
In process
Command received
Command completed
Command received