EN
11
6.
Specifications
FZ series
120
i
/
i
WP / GD
200
i
/
i
WP / GD
300
i
/
i
WP / GD
500
i
1200
i
/
i
WP / GD
2000
i
/
i
WP / GD
3000
i
/
i
WP / GD
5000
i
Weighing capacity
122g
220g
320g
520g
1220g
2200g
3200g
5200g
Maximum display
122.084g 220.084g 320.084g 520.084g 1220.84g 2200.84g 3200.84g 5200.84g
Minimum display
0.001g
0.01g
Repeatability
(Standard deviation)
0.001g
0.01g
Linearity
±
0.002g
±
0.02g
Accuracy after
calibration with
internal mass*
±
0.010g
±
0.10g
±
0.15g
*
●
Accuracy after calibration using the internal mass, with the accessory breeze break installed and in
good ambient conditions (within the temperature range of 10°C to 30°C (50°F to 86°F) with no abrupt
changes in temperature or humidity, no drafts, no vibration and no effect by magnetic fields or static
electricity).
●
With the FZ
-
i
/ FZ
-
i
WP / FZ
-
GD series balance, the environment and aging may cause internal mass
value error. Calibrate the balance using an external mass and perform maintenance regularly.
FX series
120
i
/
i
WP / GD
200
i
/
i
WP / GD
300
i
/
i
WP / GD
500
i
1200
i
/
i
WP / GD
2000
i
/
i
WP / GD
3000
i
/
i
WP / GD
5000
i
Weighing capacity
122g
220g
320g
520g
1220g
2200g
3200g
5200g
Maximum display
122.084g 220.084g 320.084g 520.084g 1220.84g 2200.84g 3200.84g 5200.84g
Minimum display
0.001g
0.01g
Repeatability
(Standard deviation)
0.001g
0.01g
Linearity
±
0.002g
±
0.02g
●
The FZ
-
i
WP / FX
-
i
WP series balances have dustproof and waterproof performances (IP65 compliant).
Содержание FZ-i Series
Страница 54: ...CT 54...
Страница 55: ...CT 55 3 2 2 ON OFF ON OFF FZ i FZ i WP FZ GD GLP GMP Title block End block pcs ct mom g 2...
Страница 56: ...CT 56 4 4 1 FZ i FZ iWP FZ GD 1 30 2 3 4 GLP output GLP 5...
Страница 58: ...CT 58 5 EC E11 3 2 During Use 3 2 EC E17 FZ i FZ i WP FZ GD EC E20 EC E21 A D EC E00 EC E01...
Страница 59: ...CT 59 EC E02 Q RE ZERO Q EC E03 1 EC E04 EC E06 EC E07...
Страница 61: ......