Display
Error code
Description
Power supply voltage fault
The voltage supplied from the AC adapter is abnormal. Check if the
AC adapter is one provided with the balance.
EC, E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC, E01
Undefined command error
An undefined command was received.
Confirm the command.
EC, E02
Not ready
A received command cannot be processed.
e.g. The balance received a Q command, but not in the weighing
mode.
e.g. The balance received a Q command while processing a
RE-ZERO command.
Adjust the delay time to transmit a command.
EC, E03
Timeout error
If the timeout parameter is set to “
”, the balance
did not receive the next character of a command within the time limit
of one second. Confirm the communication.
EC, E04
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
EC, E06
Format error
A command includes incorrect data.
e.g. The number of digits is incorrect.
e.g. The alphabet was included in the numerical value.
Confirm the command.
EC, E07
Parameter setting error
The received data exceeds the range that the balance can accept.
Confirm the parameter range of the command.
Содержание FZ-i Series
Страница 54: ...CT 54...
Страница 55: ...CT 55 3 2 2 ON OFF ON OFF FZ i FZ i WP FZ GD GLP GMP Title block End block pcs ct mom g 2...
Страница 56: ...CT 56 4 4 1 FZ i FZ iWP FZ GD 1 30 2 3 4 GLP output GLP 5...
Страница 58: ...CT 58 5 EC E11 3 2 During Use 3 2 EC E17 FZ i FZ i WP FZ GD EC E20 EC E21 A D EC E00 EC E01...
Страница 59: ...CT 59 EC E02 Q RE ZERO Q EC E03 1 EC E04 EC E06 EC E07...