EN
7
4. Calibration
4-1. Calibration Using the Internal Mass
(One-Touch Calibration, FZ
-
i
/ FZ
-
i
WP / FZ
-
GD Series Only)
This function calibrates the balance using the internal mass.
1. Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the weighing
pan.
2. Place the accessory breeze break on the balance and press the
key. The balance displays
.
3. The balance performs calibration using the internal mass. Do not allow vibration or drafts to affect the
balance.
4. After the calibration, if the "GLP output (
)" parameter of the function table is set, the balance
outputs a sensitivity adjustment report.
5. The balance will automatically return to weighing mode after calibration.
Содержание FZ-i Series
Страница 54: ...CT 54...
Страница 55: ...CT 55 3 2 2 ON OFF ON OFF FZ i FZ i WP FZ GD GLP GMP Title block End block pcs ct mom g 2...
Страница 56: ...CT 56 4 4 1 FZ i FZ iWP FZ GD 1 30 2 3 4 GLP output GLP 5...
Страница 58: ...CT 58 5 EC E11 3 2 During Use 3 2 EC E17 FZ i FZ i WP FZ GD EC E20 EC E21 A D EC E00 EC E01...
Страница 59: ...CT 59 EC E02 Q RE ZERO Q EC E03 1 EC E04 EC E06 EC E07...
Страница 61: ......