5.
Error Codes
Display
Error code
Description
EC, E11
Stability error
The balance cannot stabilize due to an environmental problem.
Check around the pan. Prevent vibration, drafts, temperature
changes, static electricity and magnetic fields, from influencing the
balance. Refer to "3-2 During Use" in the instruction manual.
To return to the weighing mode, press the
key.
Out of range error
The entered value exceeds the setting range.
Enter a correct value.
EC, E17
Internal mass error (FZ
-
i
/ FZ
-
i
WP / FZ
-
GD series only)
The internal mass application mechanism does not work properly.
Perform the weighing operation from the beginning again.
EC, E20
Calibration weight error (Positive value)
The calibration weight is too heavy. Verify the calibration mass
value. Press the
key to return to the weighing mode.
EC, E21
Calibration weight error (Negative value)
The calibration weight is too light. Confirm the calibration mass
value. Press the
key to return to the weighing mode.
Overload error
A sample beyond the balance weighing capacity has been placed
on the pan. Remove the sample from the pan.
Weighing pan error
The mass value is too light. Confirm that the weighing pan is
properly installed and calibrate the balance.
Sample mass error
The balance cannot store the sample for the counting mode or for
the percent mode because it is too light. Use a larger sample.
Unit mass error
The sample unit mass for the counting mode is too light. Storing and
using it for counting will cause a counting error. Add samples to
reach the specified number and press the
key. Pressing
the
key without adding samples will shift the balance to
the counting mode. But, for accurate counting, be sure to add
samples.
Clock battery error
The clock backup battery has been depleted. Press any key and set
the time and date. The clock and calendar function work normally as
long as the AC adapter is connected to the balance. If this error
appears frequently, contact the local A&D dealer.
Содержание FZ-i Series
Страница 54: ...CT 54...
Страница 55: ...CT 55 3 2 2 ON OFF ON OFF FZ i FZ i WP FZ GD GLP GMP Title block End block pcs ct mom g 2...
Страница 56: ...CT 56 4 4 1 FZ i FZ iWP FZ GD 1 30 2 3 4 GLP output GLP 5...
Страница 58: ...CT 58 5 EC E11 3 2 During Use 3 2 EC E17 FZ i FZ i WP FZ GD EC E20 EC E21 A D EC E00 EC E01...
Страница 59: ...CT 59 EC E02 Q RE ZERO Q EC E03 1 EC E04 EC E06 EC E07...
Страница 61: ......