36
Step 13
Select the Interval drop-down menu for Period Three (3) and set the measurement
Interval for Period Three (3) to: OFF, 1, 3, 5, 10, 15, 20
,
30, 60 or 120 minutes.
Step 14
If you want to program a fourth period, go to step 15.
If you do not want to program a fourth period, go to step 17.
Step 15
Select the Time field for Period Four (4) and enter the time you want for this period;
the end period for Period Four (4) will be the Time for Period (1). Use military time to
set the Time. See examples at the end of this section.
Step 16
Select the Interval drop-down menu for Period Four (4) and set the measurement
Interval for Period Four (4) to: OFF, 1, 3, 5, 10, 15, 20, 30, 60 or 120 minutes.
Step 17
If the intervals are set correctly, click the
OK
button to confirm the programming. If
they are not correct, reset them, or go to step 18.
Step 18
To cancel the programming and close the Recorder–Setup dialog box, click the
Cancel
button.
2-7-3-2 TM-2430 Series Setup
The recorder setup for the TM-2430 series is similar to the TM-2420/TM-2421 except that it includes
two additional options for setting the BP measurement intervals as shown in Figure 22.
Figure 22: TM-2430
series
Recorder-Setup dialog box
Description of each mode is as follows:
Mode 1
07:00-21:59 The measurement is performed every 15 minutes.
22:00-06:59 The measurement is performed every 30 minutes.
Mode 2
The AUTO ON/OFF key is pressed at rising and going to bed so that the measurement
intervals are changed and the time during sleep can be distinguished on the data.
When the “S” is off, the measurement is performed every 15 minutes.
When the “S” is displayed, the measurement is performed every 30 minutes.
Mode 3
The measurement interval can be changed six times within a maximum of 24 hours. (The
monitor can store six measurement intervals (periods) in 24 hours. A period consists of a
start time and interval.)
Содержание Doctor Pro TM-2430-13
Страница 54: ...53 4 PRINTOUT EXAMPLES Figure 36 Patient information Text data ...
Страница 55: ...54 Figure 37 Summary data Full and partial Text data ...
Страница 56: ...55 Figure 38 Summary data Awake and sleep Text data ...
Страница 57: ...56 Figure 39 Blood pressure data Text data ...
Страница 58: ...57 Figure 40 Trend graph Blood pressure and double product Graphical data ...
Страница 59: ...58 Figure 41 Correlation plots upper and histogram plots lower Graphical data ...
Страница 60: ...59 Figure 42 Circadian rhythm plots upper and HBI analysis plots lower Graphical data ...
Страница 61: ...60 Figure 43 Mini report 1 ...
Страница 62: ...61 Figure 44 Mini report 2 ...
Страница 67: ...66 MEMO ...
Страница 68: ...67 MEMO ...
Страница 69: ...68 MEMO ...