![A&D AD-4212A Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/aandd/ad-4212a/ad-4212a_instruction-manual_446302073.webp)
71
18-2 Acknowledge Code and Error Codes
When the “Serial interface function (
5if
)” parameter is set to “
erCd 1
”, the balance outputs <AK>
code or error code to each command as follows:
<AK> (06h)
Acknowledge in ASCII code.
When the balance receives a command to request data and cannot process it, the balance
transmits an error code (EC, Exx).
When the balance receives a command to request data and can process it, the balance outputs
the data.
When the balance receives a command to control the balance and cannot process it, the
balance transmits an error code (EC, Exx).
When the balance receives a command to control the balance and can process it, the balance
transmits the acknowledge code.
Among commands to control the balance, the following transmit the acknowledge code both
when the balance receives the command and when the balance has accomplished the
command. If the command cannot be processed properly, the balance transmits an error code
(EC, Exx). This error can be released using the CAL command.
CAL command (Calibration command) ON command (Display ON command)
P command (Display ON/OFF command) R command (RE-ZERO command)
R command example
When a communication error has occurred due to external noise, or a parity error has occurred
due to transmission error, the balance transmits an error code. In this case, send the command
again.