60
14-4 Memory for Unit Mass in the Counting Mode
The data memory function can store 20 sets of unit mass for the counting mode. Even if the AC
adapter is removed, the data is maintained in non-volatile memory. Among the 20 sets, “
p01
” is
the memory function for the standard counting mode.
The unit mass in memory can be recalled and used for weighing.
The unit mass in memory can be recalled and changed.
Recalling the unit mass
1 Set the “Data memory (
data
)” parameter to “
1
”.
2 Set
“
Unit selection
(
Unit
)
” parameter of the function table to
pC
(counting mode). For details,
refer to “5-2 Changing the Unit”.
Note
If “
Clr
” appears blinking in the upper left of the display, delete the data in memory.
3 Press and hold the PRINT key until the balance enters the sample unit mass selection mode
as shown below.
4 Select the unit mass to be used using the following keys.
RE-ZERO key
To increase the unit mass number by one.
MODE key
To decrease the unit mass number by one.
PRINT key
To select the unit mass number to be used.
CAL key
To cancel the operation and go to step 5.
5 To re-store the unit mass or perform Automatic Counting Accuracy Improvement (ACAI) on the
re-stored unit mass, follow the procedure described in “12. COUNTING MODE (PC)”.
Notes
ACAI cannot be performed directly on the recalled unit mass.
Using “UN:mm” command, the unit mass can be recalled (mm represents 01-20, which
corresponds to P01-P20).