17. CONNECTION TO PERIPHERAL EQUIPMENT
17-1 Connection to the AD-8121B Printer
Set the following parameters to use the AD-8121B printer.
Class
Item
Factory
setting
AD-8121B
Mode 1
Mode 2
Mode 3
dout
Data output
prt
Data output mode
0
0
,
1
,
2
,
4
,
5
3
0
,
1
,
2
,
4
,
5
ap-p
Auto print polarity
0
*
*1
Not
applicable
*1
ap-b
Auto print difference
1
d-no
Data number output
0 0
0 0
,
1
5-td
Time/date output
0 0
0 0
,
1
,
2
,
3
5-id
ID number output
0 0
0 0
,
1
pUse
Data output pause
0 0
0 0
,
1
*2
at-f
Auto feed
0 0
0 0
,
1
5if
Serial
interface
bp5
Baud rate
2
2 2 2
ptpr
Data bit, parity bit
0
0 0 0
Crlf
Terminator
0
0 0 0
type
Data format
0
0 0 1
Ct5
CTS, RTS control
0
0 0 0
*1
Set the parameter when the data output mode is set to the auto print mode (
prt 1
or
prt 2
).
*2
When multiple lines are to be printed, set the parameter to
1
.
Note
AD-8121B settings
Mode
AD-8121B DIP switches
Description
Mode 1
Prints upon data receipt
Standard mode, statistical calculation mode
Mode 2
Prints using the AD-8121B DATA key or the AD-8121B
built-in timer
Standard mode, interval mode, chart mode
Mode 3
Prints upon data receipt
Dump print mode
DIP switch 3: Handling unstable data
ON = To print unstable data
OFF = Not to print unstable data
DIP switch 4: Data input specification
ON = Use the current loop
OFF = Use the RS-232C