39
DP (Dump print) format
5if type 1
This format is used when the peripheral equipment cannot receive the A&D format.
If an AD-8121B is used, set the printer to MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
The unit, consisting of three characters, follows the data.
KF format
5if type 2
This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only
communicate using this format.
This format consists of fourteen characters excluding the terminator.
This format has no header characters.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
This format outputs the unit only for a stable value.
MT format
5if type 3
A header of two characters indicates the balance condition.
The polarity sign is used only for negative data.
The weighing data uses spaces in place of the leading zeros.
The character length of this format changes dependent upon the unit
NU (numerical) format
5if type 4
This format outputs only numerical data.
This format consists of nine characters for the AD-4212A series balance or ten characters for
the AD-4212B series balance, excluding the terminator.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is used.
(AD-4212A series balance)
(AD-4212B series balance)
W
T
+ 0 . 1 2 7 8
g
CR LF
T
W T
U S
Q T
Stable header
Stable header of counting mode
Unstable header
Header
Data
Unit
Terminator
Stable value
Unstable value
Data
Unit Terminator
S 0 . 1 2 7 8 g
CR LF
S
S D
S I
Stable header
Unstable header
Overload header
Header
Data
Unit Terminator
Data Terminator
+
0 0 0 0 . 1 2 7 8
CR LF
+
+
0 0 0 . 1 2 7 8
CR LF
+
Data Terminator