5.4 Electrical Safety

Do not access the inside of any part of the system

WARNING

There is a risk of electrical shock if you attempt to access the inside of any 
part of the system. Only authorized and qualified service personnel may 
access the inside of any part of the system.

Spilled liquids

WARNING

The electrical safety of the system may be compromised if it is exposed to 
any liquids. Do not bring liquids, such as beverages, near the TRIOS system. 
Do not spill liquids, such as cleaning agent, on the TRIOS system.

Do not use a power strip

WARNING

Failure to establish a ground connection may compromise electrical safety. Do 
not connect the Battery Charger to an ordinary power strip or extension cord.

Stress on cables

CAUTION

External cables must never be subjected to pulling stress.

Condensation

CAUTION

Variations in temperature or humidity may cause water to condense inside 
the system, which can cause damage. Always let the system reach room 
temperature before you plug it into the power source.

If the system has been subject to major temperature or humidity changes, 
wait until the system has reached room temperature before plugging it  
into the power source.

If there is visible evidence of condensation, wait at least 8 hours  
before plugging it into the power source, to ensure that no internal 
condensation occurs. 

Please see operating conditions for further information.

Disconnect the battery charger before cleaning

WARNING

The electrical safety of the system may be compromised if it is exposed  
to excessive amounts of cleaning liquids. Disconnect the battery charger 
from power before cleaning and allow to dry before reconnecting.

Disconnecting from mains

IMPORTANT NOTICE

There is no power ON/OFF switch on the system; therefore, the only reliable 
means to disconnect the system from mains is to unplug the charger power 
cord. Do not position the system so that it is difficult to unplug the charger 
power cord. 

PC connection options and requirements 

CAUTION

The PC should be located at least 1.5 m away from the patient. Do not scan 
the patient and touch the PC at the same time! You must not touch any 
peripherals connected with a cable to the PC while scanning a patient.  
See the PC’s instructions for use for further details.

5.5 Eye Safety

CAUTION

The TRIOS scanner emits bright light from its TRIOS scanner tip during 
operation. The TRIOS scanner complies with EN (IEC) 62471 (Photobiological 
safety of lamps and lamp systems). However, we recommend caution when 
handling the TRIOS scanner.

A brief glimpse of the light into the eye is not dangerous, but do not stare 
into the beam or view directly with optical instruments, and do not aim the 
beam towards other people’s eyes. In general, viewing bright light sources 
(e.g. headlights), which have a high secondary exposure potential due  
to their blinding effect, may cause temporary reduction in visual acuity  
and afterimages, leading to irritation, annoyance, visual impairment,  
and even accidents, depending on the situation.

5.6 Hygiene

For instructions on how to mount the single-use sleeve see the TRIOS Sleeve instructions for use.

For instructions on cleaning and disinfection or sterilization go to section 7. 

Always wear gloves

WARNING

To maintain a clean work environment and the 
safety of the patient, wear clean examination 
gloves when:
•  Handling the TRIOS scanner tip – changing  

the tip or handling the window.

•  Handling the TRIOS protection tip.
•  Applying the TRIOS scanner for scanning 

patients.

•  Touching the TRIOS system.

Prepare the TRIOS scanner

WARNING

In order to avoid cross contamination between patients before each use 
you must: 
•  apply a newly sterilized TRIOS scanner tip to the TRIOS scanner.
•  apply a new TRIOS scanner sleeve  to cover the TRIOS scanner.
See the detailed instruction in the TRIOS Sleeve Instructions for use. 

Reprocess system after each patient

WARNING

In order to avoid cross contamination between 
patients you must before each use:
•  clean and disinfect or sterilize relevant parts  

of the TRIOS system.

•  clean and disinfect the TRIOS scanner before 

attaching the TRIOS protection tip. Do not allow 
patient contact with the TRIOS protection tip.

•  dispose the TRIOS scanner sleeve 

Clean and disinfect TRIOS scanner before mounting TRIOS protection tip

WARNING

Always clean and disinfect the TRIOS scanner before attaching the TRIOS 
protection tip. Do not allow patient contact with the TRIOS protection tip.

Scan only with TRIOS scanner tip on 

WARNING

In order to avoid cross contamination between patients and maximal 
electrical safety the scanner tip must be mounted when the scanner is 
inserted into the patient’s mouth or when the scanner comes in contact 
with the patient in any way. Mounting of the scanner tip is required  
for the intended function of the scanner.

IMPORTANT NOTICE

Before each patient, make sure any PC connected to the system and its 
peripherals are disinfected. It is recommended to use medical-grade 
peripherals (e.g., keyboards and mice) that can be cleaned and  
disinfected easily.

5.7 Safety when Scanning

CAUTION

The TRIOS scanner is a highly precise optical instrument and must  
be handled carefully. Do not drop, shake or bump the TRIOS scanner,  
as these actions could cause it to break or malfunction.  
Follow these guidelines:
•  Always return the TRIOS scanner to its mount when not in use.
•  Do not submerge the TRIOS scanner in any liquids.
•  Do not place the TRIOS scanner on heated or wet surfaces.
•  Grip the TRIOS scanner firmly during scanning and when removing/

returning it to its mount.

IMPORTANT NOTICE

The TRIOS scanner will get warm during use. This is normal but not part  
of any patient treatment.

CAUTION

The TRIOS scanner tip is fragile and contains a delicate window.  
To avoid damage to the tip and window, take care not to bump the tip  
onto the patient’s teeth or restorations.

CAUTION

When the TRIOS system is used for pediatric patients, please pay extra 
attention while scanning the patient.

Содержание L1P-1

Страница 1: ...APR 2022 3Shape TRIOS 5 Safety and Setup Guide L1P 1 L1P 1F...

Страница 2: ...system please carefully read and follow the instructions in this safety and setup guide and in the online helpsystem Pleasealwayskeepthissafetyandsetupguidenearthesystem Please follow all safety infor...

Страница 3: ...safety requirements for medical electrical equipment by combining it with other equipment is responsible for conducting appropriate inspection and testing to ensure continued safe use of the equipment...

Страница 4: ...Systembe performed byTRIOSauthorizedtechnicalservicepersonnel It is recommended to fully charge batteries for the TRIOS scanner before use CAUTION Let the system adjust to room temperature before use...

Страница 5: ...midity 10 85 non condensing Atmospheric pressure 800 1100 hPa Transport Conditions Ambient temperature range 10 50 C 14 122 F Relative humidity 10 85 Non condensing Atmospheric pressure 620 1200 hPa 4...

Страница 6: ...e b V1 3 Vrms V2 6 Vrms Note b Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 7 GHz E1 3 V m 80 MHz to 2 7 GHz Electromagnetic environment guidelines Portable and mobile RF communications equipment shoul...

Страница 7: ...Connection The wireless TRIOS scanner is intended to be used indoors in a domestic or commercial electromagnetic environment The wireless connection is intended to relay image stream dataandcontrolda...

Страница 8: ...ion in visual acuity and afterimages leading to irritation annoyance visual impairment and even accidents depending on the situation 5 6 Hygiene Forinstructionsonhowtomountthesingle usesleeveseetheTRI...

Страница 9: ...nner during patient examination WARNING The TRIOS Sleeve must only be used once to avoid cross contamination 7 Cleaning Disinfection and Sterilization The following TRIOS system items shall be cleaned...

Страница 10: ...ants Dispose of the CaviWipe Step 3 Remove gross contaminants from metal holder Take a CaviWipe Wipe the surface of the metal holder to remove any gross contaminants Dispose of the CaviWipe Step 4 Dis...

Страница 11: ...TRIOS scanner window follow these steps for optimal scan quality Step 1 Take a damp non woven swab and gently rub it on the inside and outside of the TRIOS scanner tip window as well as on the window...

Страница 12: ...10030138A April 2022 Copyright 3Shape TRIOS A S More information at 3Shape com Let s change dentistry together 3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark...

Отзывы: