4.11 Starting and Shutting Down the TRIOS System

Please follow the steps below to start TRIOS system for the first time:

Step 1:

 Turn the PC on. 

Start 3Shape software.

Step 2:

 Log on using a default TRIOS Operator.

Step 3:

 Configure and use your TRIOS system.

Follow on-screen instructions for connecting to the wireless TRIOS scanner.

When using the TRIOS system, the user is guided by messages shown in the TRIOS software. 

The TRIOS scanner has an LED ring which indicates the status of the scanner: 

A rotating blue light indicates that the TRIOS scanner 
is waiting for the TRIOS software to connect to it. 

Three blue blinks indicate that the TRIOS scanner is 
searching for a scanner application to connect to.

A constant blue light indicates that the TRIOS 
scanner is connected to the system, but no  
scanner tip is mounted. 

A rotating green light indicates that the TRIOS 
scanner is heating the TRIOS scanner tip.

A constant green light indicates that the TRIOS 
scanner is ready to scan. 

A rotating white light indicates that the TRIOS 
scanner is waiting for scan data. 

A constant white light indicates that the TRIOS 
scanner is scanning. The light intensity decreases  
if the TRIOS scanner loses registration.

Solid yellow indicates that the battery is low.

Scanner operation

The scanner has two buttons for operation; the button closests to the tip controls scanning 
operation, the button farthest from the tip controls navigation. 

Power saving

The TRIOS scanner has integrated power control, and will go into sleep mode when not used, 
unless the software is on the Scan page.

4.12 Starting TRIOS Online Help

The TRIOS online help can be started from the TRIOS software by clicking on the Help 
button and choosing User Manual.

4.13 Storage, Operating and Transport Conditions

Operating Conditions

Ambient temperature range: +15 – +26 °C (+59  – +78.8 °F)

Relative humidity: 10 – 85 % (non-condensing)

Atmospheric pressure: 800 – 1100 hPa

The TRIOS scanner should be used in an environment maintaining a 
consistent room temperature.

Storage Conditions

Ambient temperature range: -10 – +50 °C (+14 – +122 °F)

Relative humidity: 10 – 85 % (non-condensing) 

Atmospheric pressure: 800 – 1100 hPa

Transport Conditions

Ambient temperature range: -10 – +50 °C (+14 – +122 °F)

Relative humidity: 10 – 85 % (Non-condensing)

Atmospheric pressure: 620 – 1200 hPa

4.14 Technical Specifications

Light source  
specifications

Do not view directly with optical instruments, such as spectacle 
microscopes. The product complies with EN (IEC) 62471 
(Photobiological safety of lamps and lamp systems and UL-standard 
US code of federal regulations, 21 CF 1040.10 and 1040.11.

Sterilization

Sterilize by autoclave (see section 7 for autoclave specifications)

Power input

C2BP: Battery charger 

Power input: 100-240 VAC 0.6 A, 50/60 Hz 

Protection class: IP20

Insulation class Class II

L1-2 scanner. Wireless, battery powered

3Shape TB-04 Battery: 

3,6 V, 3300 mAh, 12 Wh

Wireless

The TRIOS scanner contains an IEEE 802.11ac module

Wireless  
Transmit Power

Maximum 100 mW

Wireless  
Frequency Band

2.401-2.483 GHz

5.15-5.725 GHz (the actual frequencies are dependent on local 
regulations and the configuration of the product.)

IT Network equipment

Network 
characteristics

802.11ac (5GHz)

Capacity for at least 40 Mbps per scanner.

Network 
configuration

Bi-directional traffic permitted between TRIOS scanner and PC.

mDNS enabled (required for scanner discovery).

Redirection of traffic must be disabled.

Channel width set to at least VHT40 (40MHz).

Ports used:
•  TCP ports (Scanner): 20-21, 80 and 23795
•  UDP Ports (Scanner): 58218 and 58220-58230
•  UDP Ports (PC and Router): 67 (DHCP), 5353 (mDNS)

Security 
specification

WPA2-PSK authentication

IMPORTANT NOTICE

If you connect your TRIOS intraoral scanner system to your IT Network 
equipment, it is strongly recommended to apply network security 
measures and to analyse, evaluate, and control the IT network.

4.15 Electromagnetic Compability

The TRIOS system has been tested according Medical electrical equipment – Part 1-2: 
General requirements for basic safety and essential performance – Collateral Standard: 
Electromagnetic disturbances – Requirements and tests, EN/(IEC) 60601-1-2 ed. 4.0, 2015.

WARNING

Medical electrical equipment needs special precautions regarding EMC 
and needs to be installed and put into service according to the EMC 
information provided in this document.

IMPORTANT NOTICE

Portable and mobile RF communications equipment can affect medical 
electrical equipment.

Содержание L1P-1

Страница 1: ...APR 2022 3Shape TRIOS 5 Safety and Setup Guide L1P 1 L1P 1F...

Страница 2: ...system please carefully read and follow the instructions in this safety and setup guide and in the online helpsystem Pleasealwayskeepthissafetyandsetupguidenearthesystem Please follow all safety infor...

Страница 3: ...safety requirements for medical electrical equipment by combining it with other equipment is responsible for conducting appropriate inspection and testing to ensure continued safe use of the equipment...

Страница 4: ...Systembe performed byTRIOSauthorizedtechnicalservicepersonnel It is recommended to fully charge batteries for the TRIOS scanner before use CAUTION Let the system adjust to room temperature before use...

Страница 5: ...midity 10 85 non condensing Atmospheric pressure 800 1100 hPa Transport Conditions Ambient temperature range 10 50 C 14 122 F Relative humidity 10 85 Non condensing Atmospheric pressure 620 1200 hPa 4...

Страница 6: ...e b V1 3 Vrms V2 6 Vrms Note b Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 7 GHz E1 3 V m 80 MHz to 2 7 GHz Electromagnetic environment guidelines Portable and mobile RF communications equipment shoul...

Страница 7: ...Connection The wireless TRIOS scanner is intended to be used indoors in a domestic or commercial electromagnetic environment The wireless connection is intended to relay image stream dataandcontrolda...

Страница 8: ...ion in visual acuity and afterimages leading to irritation annoyance visual impairment and even accidents depending on the situation 5 6 Hygiene Forinstructionsonhowtomountthesingle usesleeveseetheTRI...

Страница 9: ...nner during patient examination WARNING The TRIOS Sleeve must only be used once to avoid cross contamination 7 Cleaning Disinfection and Sterilization The following TRIOS system items shall be cleaned...

Страница 10: ...ants Dispose of the CaviWipe Step 3 Remove gross contaminants from metal holder Take a CaviWipe Wipe the surface of the metal holder to remove any gross contaminants Dispose of the CaviWipe Step 4 Dis...

Страница 11: ...TRIOS scanner window follow these steps for optimal scan quality Step 1 Take a damp non woven swab and gently rub it on the inside and outside of the TRIOS scanner tip window as well as on the window...

Страница 12: ...10030138A April 2022 Copyright 3Shape TRIOS A S More information at 3Shape com Let s change dentistry together 3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark...

Отзывы: