EO Abator Operation and Installation Manual

2

Read, understand, and follow all safety information contained in these instructions prior to using the 3M EO Abator. 

Retain these instructions for future reference.

Intended Use:

The 3M™ EO Abator is intended to remove ethylene oxide (EO) from the exhaust of 3M™ Steri-Vac™ Sterilizers and 

Aerators. The EO Abator converts EO into carbon dioxide and water vapor through a heated catalytic process. Use with 

other than 3M™ Steri-Vac™ Sterilizers and Aerators could lead to an unsafe condition.

Explanation of Signal Words and Symbols

Symbol Title

Symbol

Description and Reference

Caution

Indicates the need for the user to consult the instructions for use for important 

cautionary information such as warnings and precautions that cannot, for a variety 

of reasons, be presented on the medical device itself. Source: ISO 15223, 5.4.4

Flammable 

Product is classified as flammable. Source: Globally Harmonized Symbols for 

Classifying Hazardous Chemicals 

Health hazard

The product contains an ingredient classified as a health hazard. Source: Globally 

Harmonized Symbols for Classifying Hazardous Chemicals

Warning, electricity

To warn user to avoid coming into contact with electricity. Source: ISO 7010-W012

Follow instructions 

for use

To signify that the instructions for use must be followed. Source: ISO 7010-M002

Caution, hot 

surface

Indicates product can be hot and should not be touched without taking care. 

Source: IEC 60417-5041

For more information see, HCBGregulatory.3M.com

 WARNING

 

To reduce the risk associated with hazardous voltage, which if not avoided, could result in death or 

serious injury:

-  Do not attempt to service and/or open the EO Abator.
-  Service should only be performed by 3M Authorized Service Personnel.

• 

To reduce the risk associated with high temperature, which if not avoided, could result in death or serious 

injury and/or property damage:

-  Do not touch the exhaust lines or the immediate area around them.
-  Do not place items next to the EO Abator. Follow clearance requirements listed in the Operation and 

Installation Manual.

Содержание 50AN

Страница 1: ...3 Operation and Installation Manual Models 50AN 50AE 50AJ EO Abator ...

Страница 2: ...uctions 8 Planning the EO Abator Installation 8 Service Installation Review and Checkout 8 Optional Configurations 8 Purchaser s Responsibility 10 EO Abator Efficiency Testing 10 Unpacking the EO Abator 11 Locating the EO Abator System 11 Section 3 Installation Requirements 12 General Electrical Requirements 12 Electrical Requirements Models 5XL 8XL and GS Series Sterilizers 12 EO Inlet Line Requi...

Страница 3: ...ed Symbols for Classifying Hazardous Chemicals Health hazard The product contains an ingredient classified as a health hazard Source Globally Harmonized Symbols for Classifying Hazardous Chemicals Warning electricity To warn user to avoid coming into contact with electricity Source ISO 7010 W012 Follow instructions for use To signify that the instructions for use must be followed Source ISO 7010 M...

Страница 4: ...Remove contaminated clothing and wash before reuse Discard all leather items exposed to EO If symptoms develop seek medical attention Eye Contact Immediately flush with large amounts of water for at least 15 minutes Remove contact lenses if easy to do so Promptly seek medical attention If Swallowed Rinse mouth DO NOT INDUCE VOMITING Immediately seek medical attention Leak or Fire In case of leak o...

Страница 5: ...f Conformity EMC Compliance The Model 50AN and Model 50AE EO Abators comply with IEC EN 61326 1 as confirmed in the Certificate of Compliance generated by 3M The Model 50AN and Model 50AE EO Abators comply with the EMC requirements of the CE mark EMC Directive 2014 30 EU The Model 50AN Abator complies with the Australian EMC requirements as confirmed in the Supplier s Declaration of Conformity tha...

Страница 6: ...rvice Area Top 900 mm 36 in Left Side 500 mm 20 in Right Side 500 mm 20 in Front 500 mm 20 in Back 100 mm 4 in Heat Source 6 kW Electric Air Duct Heater Minimum Air Flow 1 4 NCMM 50 SCFM Weight 163 kg 360 lbs Maximum EO Feed Rate 7 7 g min 0 017 lbs min Exhaust Temperature Idle Ready 138ºC 280ºF Typical Operating Processing EO 238ºC 460ºF Upper Limit 260ºC 500ºF ...

Страница 7: ...properly installed with 3M Steri Vac Sterilizers and or Aerators the temperature should not exceed 260ºC 500ºF If this value is exceeded damage to the catalytic cell and other internal components will result The EO Abator achieves at least 99 9 efficiency time weighted average in removing EO during sterilization portion of cycle when EO 100ppm at normal operating temperatures and concentrations Th...

Страница 8: ...ould be shut down When shutting down the Abator make sure the sterilizer is not venting EO and that the Abator is not processing EO In an emergency situation the Abator can be shut down at any time To shut down the system in an emergency have an authorized person switch the external disconnect and main circuit breaker in the control panel to the off position Typical Abator System Cycle for 5XL 8XL...

Страница 9: ...ator unit The GS Series sterilizer must be connected to a separate Abator The Model 5XL or Model 8XL sterilizers CANNOT be combined with a GS Series sterilizer when connecting to an Abator Up to two additional Model XL Aerators can be added to the installation The EO Abator interfaces with the Model 5XL Model 8XL and GS Series sterilizers to start and stop automatically For 3M Model XL Aerator ins...

Страница 10: ...ontrol panel connection AWG 16 Are jumper wires installed if only one sterilizer is connected to the EO Abator system Yes No EO Inlet Vent line from Sterilizer to Abator 17 Does the EO inlet line connect directly to the EO Abator system without being terminated into any other exhaust system Yes No 18 What type of material is the EO inlet line made of 19 What is the diameter in mm in of the EO inle...

Страница 11: ...h its labeling Injury to persons or property can result unless the operating instructions are followed carefully It is the purchaser s responsibility to provide the necessary machine service requirements to the area where the sterilizer and EO Abator system are to be installed These services include electricity EO inlet line safety vent line and a dedicated exhaust system Because of varying local ...

Страница 12: ... operation The EO Abator system exterior cabinet can reach temperatures as high as 50 C 122 F Temperatures within the machine can approach 260 C 500 F Select an appropriate site well in advance of purchasing the EO Abator system Contact your building engineering department for help in selecting a site It is suggested that the EO Abator system be installed indoors This saves added installation expe...

Страница 13: ... case is the indicator panel to be mounted more than a maximum of 40 M 130 ft away from the EO Abator system The indicator panel must be located in the sterilizer area A ground wire from the ground lug in the control panel to earth ground is required Use the same wire criteria as listed above for the indicator panel to determine the proper wire AWG for the distance run between the control panel an...

Страница 14: ...on refer to the appropriate XL Series or GS Series 3M Steri Vac Sterilizer Site Planning and Installation Guide Exhaust System Requirements The temperature of the air exhaust from the EO Abator can be as high as 260 C 500 F The EO Abator exhaust can contain low concentrations of EO typically less than 2 5 PPM The EO Abator exhaust must be connected to a dedicated exhaust system supplied by the cus...

Страница 15: ...e used for runs up to 183 meters 600 ft long minus 12 meters 40 ft for each 90 elbow The Metalbestos adapter part number 78 8078 9813 1 is needed to connect the EO Abator exhaust to the Metalbestos ducting and is available through the 3M Health Care Service Center Connect the Metalbestos adapter to the EO Abator exhaust using high temperature RTV between the two flanges and then clamp the flanges ...

Страница 16: ...EO Abator Operation and Installation Manual 15 Section 4 Figures Figure 1 Negative Pressure System Arrangement 1 Positive Pressure System Arrangement 2 ...

Страница 17: ...EO Abator Operation and Installation Manual 16 Figure 2 View of Inside the EO Abator Control Box Model 50AN Connect 220 230 VAC to circuit breaker 1 and connect ground wire to ground lug 2 ...

Страница 18: ...EO Abator Operation and Installation Manual 17 Figure 2 cont Model 50AE Connect Three Phase 400 VAC and Neutral to circuit breaker 1 and connect ground wire to ground lug 2 ...

Страница 19: ...EO Abator Operation and Installation Manual 18 Figure 2 cont Model 50AJ Connect Three Phase 200 VAC to circuit breaker 1 and connect ground wire to ground lug 2 ...

Страница 20: ...EO Abator Operation and Installation Manual 19 Figure 3 For 5XL 8XL and GS Series Sterilizers D Connector ...

Страница 21: ... terminal 28 to 26 If a second sterilizer is connected connect wires to these terminals 8XL 5XL D Connector Sterilizer 2 PIN 4XL 5XL 3M Steri Vac Sterilizer 2 Wire to EO Abator System Indicator Panel Terminal 10 20 15 15 22 22 22 26 13 25 25 4 38 16 14 40 21 Note One 5XL 8XL GS Series D connector cable part number 78 8078 6209 5 is provided with each EO Abator Additional cables are available throu...

Страница 22: ...EO Abator Operation and Installation Manual 21 Figure 5 Clearances Required ...

Страница 23: ...EO Abator Operation and Installation Manual 22 Figure 6 Multiple Hoods Aerators and EO Abator Vented into a Common System Note Safety Vent Line required but not shown ...

Страница 24: ...EO Abator Operation and Installation Manual 23 Figure 7 Aerator Venting Connection Aerator Electrical Connection ...

Страница 25: ...at is familiar with the testing and handling of EO 3M does not perform efficiency testing For servicing information in the USA contact your local 3M service representative or the 3M Health Care Service Center at the following Address 3M Health Care Service Center Bldg 502 1W 01 3350 Granada Ave North Suite 200 Oakdale MN 55128 If you have questions call 3M Field Service 1 800 688 5888 Outside the ...

Страница 26: ...3M is a trademark of 3M Used under license in Canada 2020 3M All rights reserved 3M est une marque de commerce de 3M utilisées sous licence au Canada 2020 3M Tous droits réservés Manual Installation 05 264058_3 Issue Date 2019 12 rev 3 34 8725 3776 5 3M Deutschland GMbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany ...

Отзывы: