Display Readings

Normal Operating Mode

The LCD will display 0 ppm for gas levels below  ppm. Gas levels of  ppm and higher will be displayed in 

increments of  ppm.

Performing Self-test

A self-test can be performed at any time by pressing the On/Off/Menu button once from the normal operating 

mode. This will test the battery, electronic circuitry and alarm functions to ensure they are working properly. 

The self-test does not test the performance of the sensor. This must be done through a calibration or 

calibration verification test. See Calibration Verification and Calibration section of these 

User Instructions. 

The self-test will be indicated by the “test” icon being displayed on the LCD for about  seconds (Fig. ). The 

buzzer, LED and vibrator will activate and then “CAL” will be displayed for  seconds (Fig. 4). If the button 

is pressed within  seconds of “CAL” being displayed, the unit will enter calibration mode (see Calibration 

Verification Test and Calibration section). After successful completion of a self-test, the “check” icon (Fig. ) 

will appear on the display for 4 hours, and the unit will be in normal operating mode. 
If a self-test fails, the LCD will display “Err” (Fig. 6). Press button to clear. 

Do not use the instrument until 

the reason for the “Err” message has been determined and corrected.

Viewing Peak Levels

From normal mode, press the button twice and the peak reading will be displayed. 

 

WARNING

To clear the peak reading, wait  seconds; “Clr” will be displayed (Fig.7). Press 

the button once to clear and the value will be set to zero so long as the unit is in a 

non-contaminated environment. The instrument will go back to normal operating 

mode after  seconds.

Calibration Reminder Function

The unit has a calibration reminder function that can be enabled and disabled 

by the user (factory default is enabled). When enabled, the “TEST” icon will flash 

when the time since last calibration exceeds 0 days (Fig. 8).
In the event a user has a fixed calibration schedule less than every 0 days, the 

calibration reminder function may be disabled. To disable this function: press the 

 

WARNING

Fig. 7

Fig. 

Each time the unit is turned on, it performs a self-test, which activates the audible, vibratory and visual 

alarms. If the self-test fails, or all the alarms do not activate, do not use. Failure to do so may adversely 

affect product performance and 

result in sickness or death.

Fig. 

Fig. 4

Fig. 

Fig. 6

Fig. 8

Alarms

Viewing Alarm Set Points

To display alarm set points, press the button 4 times from the normal operating mode. The alarm will be 

displayed. Press the button again to display the TWA alarm and STEL alarm, respectively.

Setting Alarm Levels

The units factory alarm level default set point is  ppm. The alarm level can be changed to one of four 

preset values, depending on user needs and applications. The four user selectable alarm settings are: ppm, 

ppm, 0ppm, 00ppm.
To change the alarm value, press the button 4 times from the normal operating mode. Wait  seconds until 

the display flashes, then press the button repeatedly until the desired alarm point is shown. Wait  seconds 

until the display returns to the normal mode. The new alarm point is now set.

button  times, the “Off/On” status will be displayed, wait  seconds until the “TEST” icon begins to flash. To 

re-enable this function: press the button  times, the off/on status will be displayed, wait  seconds until the 

“TEST” icon begins to flash.
Press the button to toggle between off/on. After  seconds it will return to normal mode.

Fig. 9

Alarm

An alarm condition is indicated by a slow alarm sequence of buzzer, red LED, 

“ALARM” display and vibrator actuated every . seconds (Fig. 9).

TWA and STEL Alarm

When a TWA or STEL alarm point is reached, the alarm is actuated. The “tA” is 

displayed in the digital display area alternating every second with the gas level 

when the TWA alarm is actuated. The “StL” is displayed in the digital display area 

alternating every second with the gas level when the STEL alarm is actuated.
To clear TWA, from the normal operating mode, press the button 6 times (Fig. 

0). The display will toggle between “tA” and the TWA alarm level setting. Wait  

seconds until the display shows “Clr,” then press the button once (Fig. ).
To clear STEL, from the normal operating mode, press the button 7 times (Fig. 

). The display will toggle between “StL” and the STEL alarm level setting. Wait  

seconds until the display shows “Clr,” then press the button once (Fig. ).

Immediately exit any environment that causes an alarm condition on the monitor. Failure to do so may 

result in sickness or death.

Do not cover or obstruct display, audible alarm opening or visual alarm cover. Doing so may adversely 

affect product performance and 

result in sickness or death.

Vibrator and LCD may not function effectively below –4 ºF (–0 ºC). Using the instrument below this 

temperature may adversely affect product performance and 

result in sickness or death.

Fig. 0

Fig. 

Fig. 

Содержание 110XLS Series

Страница 1: ...eur de gaz unique 3M de s rie 110XLS monoxyde de carbone CO Conserver ces directives titre de r f rence Monitor de un Solo Gas Serie 110XLS Instrucciones de uso para el Monitor de un solo gas 3M Serie...

Страница 2: ...d result in over exposure and cause sickness or death For proper use see supervisor or User Instructions or call 3M in U S A 1 800 243 4630 In Canada call Technical Service at 1 800 267 4414 Use only...

Страница 3: ...tors the ambient air that enters the sensor through the sensor inlet opening by the process of passive diffusion If the level of the target gas detected by the sensor reaches a factory preset alarm po...

Страница 4: ...Monoxide CO It is not to be considered all inclusive nor is it intended to replace the policy and procedures for each facility If you have any doubts about the applicability of the equipment to your...

Страница 5: ...ckness or death Fig 2 Fig 4 Fig 5 Fig 6 Fig 8 Alarms Viewing Alarm Set Points To display alarm set points press the button 4 times from the normal operating mode The alarm will be displayed Press the...

Страница 6: ...g a calibration or calibration verification test bump test only use certified calibration gas at the required concentration level Do not calibrate with expired calibration gas A calibration verificati...

Страница 7: ...ation Kit 17L 10 LEL Methane 35 PPM CO 18 Vol OX with Model 713 Fixed Flow 0 2 LPM Regulator and Case 1 Case 933 000 021 3M 4 Gas Mixture Methane Calibration Kit 58L 10 LEL Methane 35 PPM CO 10 PPM HS...

Страница 8: ...of a regular maintenance program Any implied warranties arising out of the sale of 3M s Portable Gas Detectors or Monitors including but not limited to the implied warranties of merchantability and fi...

Страница 9: ...ATION 22 Mise en marche et arr t de l appareil 23 Lectures de l affichage 23 Mode d utilisation normale 23 Effectuer un auto test 23 Affichage des concentrations maximales 24 Fonction de rappel d talo...

Страница 10: ...hage l ouverture de l alarme audible ou le panneau de l alarme visuelle S assurer que l orifice d admission du capteur est d gag et exempt de d bris S assurer que le dispositif d talonnage est retir a...

Страница 11: ...able en usine S curit intrins que CSA U classe 1 division I groupes A B C et D T4A Alimentation Pile au lithium de 3 6 volts non rechargeable install e en permanence rempla able en usine Dur e utile d...

Страница 12: ...t Pour tout renseignement sur l utilisation ad quate de ce produit consulter son superviseur ou lire les Directives d utilisation ou communiquer au Canada avec le Service technique au 1 800 267 4414 S...

Страница 13: ...hage revienne au mode normal La nouvelle valeur de l alarme est maintenant r gl e Alarme de d passement de plage Une alarme de d passement de plage est signal e par le clignotement l affichage de la v...

Страница 14: ...probl me n a pas t d termin et corrig Ne pas couvrir ni obstruer l affichage l ouverture de l alarme audible ou le panneau de l alarme visuelle S assurer que l orifice d admission du capteur est d gag...

Страница 15: ...intrins que ce qui peut nuire au bon fonctionnement du produit et provoquer des probl mes de sant ou la mort MISE EN GARDE Trousses d talonnage comprenant le gaz indiqu le d tendeur et l tui Pi ce n...

Страница 16: ...s OX EX HS et CO 2 ans compter de la date de fabrication Ensembles de capteurs de la s rie 740 2 ans compter de la date de fabrication Tous les autres capteurs 1 an compter de la date de fabrication C...

Страница 17: ...e pantalla 39 Modo de operaci n normal 39 Realizaci n de prueba de auto verificaci n 39 Vista de niveles pico 40 Funci n para recordatorio de calibraci n 40 Alarmas 40 Vista de puntos establecidos de...

Страница 18: ...ras incorrectas Lo anterior significa que los niveles presentes de gas monitoreado pueden ser mayores y provocar sobre exposici n lo cual pudiera ocasionar alguna enfermedad o incluso la muerte del us...

Страница 19: ...te a 450 MHz 5 watts de radio 61 cm 2 pies Teclado Operaci n de un bot n Rango de temperatura de operaci n Ver la tabla a continuaci n Exactitud de sensor despu s de calibraci n Mon xido de Carbono CO...

Страница 20: ...e la secci n Calibraci n y verificaci n de calibraci n Despu s de haber completado con xito una prueba de auto verificaci n en la pantalla aparecer el icono check Fig 5 por 24 horas y el equipo cambia...

Страница 21: ...antalla regrese al modo normal Se ha activado el nuevo nivel de alarma ADVERTENCIA Abandone de inmediato cualquier ambiente que active la alarma del monitor No hacerlo puede causar enfermedad o muerte...

Страница 22: ...ado en el instrumento Conecte la manguera del regulador de gas del contenedor de gas para calibraci n a la campana para calibraci n Fig 15 Abra la v lvula de control de flujo de gas Compare los valore...

Страница 23: ...l producto y resultar en enfermedad o muerte 933 000 004 Kit para calibraci n 3M mezcla de 4 gases Pentano 58L 10 LEL Pentano 35 PPM CO 10 PPM HS 18 Vol OX con Regulador modelo 715 de flujo fijo 0 2 L...

Страница 24: ...a impl cita de que los equipos son similares a los que se ofrecen en el mercado o garant a de que son id neos para cualesquier otro fin diverso al se alado en este Manual por lo que el usuario es el r...

Страница 25: ...ade 55 Leituras do Display Digital 55 Modo de Opera o Normal 55 Realizando um Auto Teste 55 Visualizando N veis de Pico 56 Fun o de Lembrete de Calibra o 56 Alarmes 56 Visualizando os Pontos de Ajuste...

Страница 26: ...bump test antes de cada uso Se o instrumento n o puder ser calibrado n o o utilize at que a causa possa ser determinada e corrigida CUIDADO Este instrumento cont m uma bateria de l tio Descarte de ac...

Страница 27: ...a Intr nseca CSA EUA Classe I Divis o 1 Grupo A B C D T4A Energia Bateria de l tio de 3 6 volts n o recarreg vel permanentemente montada substitu vel pelo fabricante Vida til da Bateria 2 anos a part...

Страница 28: ...sensor Isto deve ser feito atrav s de uma calibra o ou um teste de verifica o da calibra o Consulte a se o Calibra o e Teste de Verifica o de Calibra o destas Instru es de Uso O auto teste ser indicad...

Страница 29: ...o 4 vezes a partir do modo de opera o normal Espere 5 segundos at que o display pisque e ent o pressione o bot o repetidamente at que o n vel de alarme desejado seja exibido Espere 5 segundos at que...

Страница 30: ...ando o display indicar GAS alternando com o n vel de g s esperado Fig 20 Aplique o g s apropriado 35 ppm para CO dentro de 30 segundos Uma vez que o g s detectado o display indicar SPn alternando com...

Страница 31: ...4 Kit de Calibra o 3M Mistura de 3 Gases Metano 34L 10 LEL Metano 35 PPM CO 18 Vol OX com Regulador de Fluxo 0 2 LPM Modelo 713 e Estojo 1 Caixa 933 000 202 Kit de Calibra o 3M Mistura de 4 Gases Pent...

Страница 32: ...ecidos acima A 3M n o ser respons vel pela perda do uso de quaisquer de seus produtos ou por custos ou despesas incidentais ou conseq entes ou por danos sofridos pelo comprador ou por qualquer outro u...

Страница 33: ...4T1 Division des produits d hygi ne industrielle et de s curit environnementale de 3M Compagnie 3M Canada C P 5757 London Ontario N6A 4T1 3M M xico S A de C V Av Santa Fe No 190 Col Santa Fe Del lvaro...

Отзывы: