-18-
10.2 Viewing Test data and Results using a Computer
10.2.1 Downloading Winfilter Software to a Computer
In order to view program settings, data and results on a computer, the Winfilter 3.01 version software needs to be installed on the computer. For in-
stallation, place the provided CD disk containing the Winfilter 3.01 and all support software in the computer CD drive, follow all instructions provided
on the computer during the installation process. Upon completion of the program installation, a Winfilter 3.0 icon will appear on computer desktop.
10.2.2 Transferring Test Data and Results to a Computer
Using provided USB (A Type) & USB mini-b cable, connect the 101-Series Integrity Test device to a computer with the installed Winfilter 3.01 data
backup program. Turn on the 101-Series Integrity Test device. Access the Winfilter program by either clicking on the computer desktop Winfilter
icon (or the Windows icon accessing computer program directories). The Winfilter3.01 program allows you to program, transfer, view and print test
program parameters, test data and results.
Notes:
If you have carried out a data backup, all data in the device are preserved, but are marked therein as backed up. Thus, they are approved for being
overwritten with new data, without requiring the appearance of a warning message. The memory is thus once again freely available.
Info -
Data Loss!
Perform the data backup as soon as possible, when you receive a warning message from the device that only ten, or fewer than ten, free memory
locations are available. If the memory is full, the device starts to overwrite the oldest data with new measurements. Thus, the oldest data is gradu-
ally lost.
Info
11. How to run Device Auto-Test
The 3M™ 101 Series Integrity Test Device performs constant monitoring of the most important operating data. If an error arises, it is reported
immediately.
The device displays important error messages, such as leaks in the filter, until the user confirms them by pressing any key. Non-critical error
messages are automatically hidden after eliminating the cause of the error.
Based on the following table, you can isolate the possible errors that arise, along with the possible causes. If only specialized personnel can
resolve an error that arises, this is noted in the “Remedy” column: “Contact 3M Purification Customer Service”.
Error
Cause
Remedy
Warning message:
Battery pack is nearly empty
Charge battery pack
Battery pack nearly empty
Error message:
Charge battery
Battery pack is empty
Connect power supply and charge battery
pack
The battery is running out quickly
Battery pack is aged
Change battery pack
Error message:
Overpressure
The pressure is greater than the specified
measurement range! As it increases further,
the device may be damaged.
Reduce applied pressure
Calibration parameters, measurement error
In the Service Menu:
1. Check calibration parameters
2. Contact 3M Purification Customer Service
Pressure sensor not operating correctly
Contact 3M Purification Customer Service