9.4 Entering Test Program
Programming may be performed manually on the device, or with the assistance of the WinFilter3.01 software. Programming through WinFilter3.01
is particularly much easier regarding the entry of model numbers and series numbers, and is therefore recommended.
Notes:
•
The entry arrow appears bright if an entry is not in the valid range. As soon as the entry is in the valid range, the arrow appears black.
•
Entries may be deleted with the “#” hash key.
-15-
Start the TEST PROGRAMS function.
Enter the desired program number between 1 and 19, e.g. 3.
The program number is displayed .
Enter the parameters shown in the following. Note the limits provided in the above table.
A filled triangle (
X
) in front of the value indicates that the value is within the valid range. Otherwise, an
empty triangle is displayed.
Using the # key will delete a value.
Enter the diffusion rate per filter cartridge. If filters are larger than 10" is used, multiply the diffusion rate by
the number of 10" modules per filter.
Enter the maximum value according to your test cartridge product brochure.
If different cartridge sizes are contained in the filter, you can also indicate the total value of the diffusion rate
and only one cartridge here.
Indicate how many cartridges are included in the housing to be tested.
Enter the test pressure according to the data sheet for your filter.
Enter the stabilization period.
Enter the test period.
TEST PROGRAMS ?
PROCEED >
TEST PROGRAM NO. 1
PROCEED >
TEST PROGRAM NO. 3
PROCEED >
START
START
PORE SIZE
X
0.33 µm >
DIFFUSION RATE
PER CARTRIDGE >
PER CARTRIDGE
MAX
X
20.0ml/min
PER CARTRIDGE
MAX
X
20.0ml/min
TEST PRESSURE
X
12.00 psi >
STABILIZATION TIME
X
180sec >
TEST TIME
X
300sec >