6
3.
Remove the 4 screws using a Philips #2 screwdriver and remove the rear cover.
4.
Push the door inwards and upwards to gain a decent handhold on the inside of the front cover
(
1
), then pull outwards and upwards to remove the front cover from the rest of the unit (
2
).
5.
Turn the front cover over to gain access to the cables.
1
2
Содержание Fire CR dental
Страница 1: ...Service Manual Hardware revisions AH and later ...
Страница 2: ...1 ...
Страница 9: ...8 8 Set the reader back upright and carefully lift the top cover off ...
Страница 10: ...9 9 Remove the 4 screws fastening the inner chassis using a 2 5 mm hex key and lift it off ...
Страница 13: ...12 4 Remove the 3 screws using a 2 5 mm hex key 5 Replace the front board and install in reverse order ...
Страница 19: ...18 13 Replace the optics bracket and install in reverse order 1 2 ...
Страница 39: ...38 7 Replace the bottom board and install in reverse order ...
Страница 63: ...62 ...