56
Scanner calibration
To calibrate the FireCR Dental you will need an IP size 4C, a generator and preferably a dosimeter.
Before calibrating ensure that the unit is scanning with
HD
resolution by looking in the top of the LCD
display on the unit.
If the unit is scanning with
SD
resolution, enter the settings menu by pressing on the cog wheel on the
LCD display of the unit.
Select the
Resolution
field.