![3Disc Fire CR dental Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/3disc/fire-cr-dental/fire-cr-dental_service-manual_442187049.webp)
48
4.
Loosen the 2 socket screws from the step motor drive shaft using a 1.5 mm hex key, you may
have to rotate the step motor to gain access to the screws.
5.
Lift the stepper motor up.
6.
Remove the 4 screws fastening the motor mount to the stepper motor and reuse the motor
mount during installation.
7.
Install in reverse order.
Содержание Fire CR dental
Страница 1: ...Service Manual Hardware revisions AH and later ...
Страница 2: ...1 ...
Страница 9: ...8 8 Set the reader back upright and carefully lift the top cover off ...
Страница 10: ...9 9 Remove the 4 screws fastening the inner chassis using a 2 5 mm hex key and lift it off ...
Страница 13: ...12 4 Remove the 3 screws using a 2 5 mm hex key 5 Replace the front board and install in reverse order ...
Страница 19: ...18 13 Replace the optics bracket and install in reverse order 1 2 ...
Страница 39: ...38 7 Replace the bottom board and install in reverse order ...
Страница 63: ...62 ...