PMT gain verification and adjustment
1.
Expose the IP with
1100µGy
, position it on the tray and close it to scan on the low scan
position.
2.
Compare the readout value to the tolerance for the high dose calibration.
If the readout value is within the tolerances for the high dose calibration, continue onto
Scanner calibration
.
If the readout value is outside the high dose calibration tolerances, open the scanner
control panel and check the box next to the “Edit Params” label.
High dose tolerance.
Low dose scan
readout value.
Edit parameters
checkbox