Instruction Manual

USS-DBS86 Precision Balance

Summary of Contents for USS-DBS86

Page 1: ...Instruction Manual USS DBS86 Precision Balance...

Page 2: ...ety Precautions 1 Design and Function 3 2 1 Components 3 2 2 Keyboard 4 2 3 Display Panel 5 Unpacking 6 Before Using 7 Operation 8 5 1 Basic Weighing 8 5 2 Unit Switching 9 5 3 Counting 9 Setting 11 6...

Page 3: ...cessories Please read the manual completely and follow the usage instructions before installation and operation as this will help you to make full use of the functions and performance of the USS DBS86...

Page 4: ...pecification of the balance If excessive voltage is used the balance may overheat and be damaged or cause a fire Operate the balance on a stable rigid and flat table Handle the balance carefully It is...

Page 5: ...3 Design and Function 2 1 Components 1 Housing 2 Housing 3 Leveling foot 4 Battery Cover 5 Dust Cover 6 Weighing Pan...

Page 6: ...n the power on or off Enter the counting function Long press with TARE to enter the minimum weight setting Tare the balance Long press with SET to enter the minimum weight setting Performs weight cali...

Page 7: ...5 2 3 Display Panel No Description 1 Indicates positive values Indicates negative values Stabilization indicator Disappearance indicates stabilization of balance 2 Weight value 3 Weight unit...

Page 8: ...ness The following accessories are part of the standard equipment for your new USS DBS86 Precision Balance 1 Balance Main Body 1 Weighing Pan 1 Power Cable 1 Instruction Manual 1 Quality Certification...

Page 9: ...dicator is centered Accurate horizontal positioning is necessary for repeatable measurements and exact results To compensate for small irregularities or tilts at this location the instrument needs to...

Page 10: ...eighing a sample that needs to be in a container 1 Open the glass door of the weighing chamber and place the weighing container on the weighing pan then close the glass door again 2 Wait for the displ...

Page 11: ...ing function that meets various industrial counting requirements especially the counting function for small components To ensure the accuracy of the counting function for small components it is necess...

Page 12: ...10 PCS again displayed and the corresponding number appears Exit counting function 1 Press SET and will be displayed 2 After that ERR will be flashing displayed 3 Press SET to exit the counting funct...

Page 13: ...will be displayed as 0 0g when the sample is less than 0 3g Therefore when using this balance for weighing samples less than 0 3g the user can set the minimum weighing to 0 1g 1 Long press TARE and SE...

Page 14: ...apacity 2100 g 3100 g 5100 g 6100 g 10000 g Readability 0 1 g Repeatability 0 2 g Linearity Deviation 0 3 g Stabilization Time Within 3 seconds Pan Size 6 3 x 6 3 inches 160 x 160 mm Interface RS232 P...

Page 15: ...13...

Page 16: ...Contact Feel free to visit our website www ussolid com You can email us at service ussolid com You can call one of our friendly customer service representatives at 1 800 209 4177...