background image

Instruction Manual

USS-DBS93(94)

Waterproof Bench Scale

Summary of Contents for USS-DBS93

Page 1: ...Instruction Manual USS DBS93 94 Waterproof Bench Scale ...

Page 2: ...mponents 3 2 2 Keyboard 4 2 3 Display Panel 5 Unpacking 6 Before Using 7 4 1 Leveling the scale 7 4 2 Calibration 9 Operation 10 5 1 Basic Weighing 10 5 2 Checkweighing 11 Setting 13 6 1 Unit Switching 13 Maintenance 15 7 1 Cleaning 15 7 2 Disposal 15 Troubleshooting 16 Technical Data 17 ...

Page 3: ...ther samples with water U S Solid s dedicated customer service staff are available to answer any inquiries regarding applications and accessories Please read the manual completely and follow the usage instructions before installation and operation as this will help you to make full use of the functions and performance of the USS DBS93 94 Waterproof Bench Scale 1 1 Safety Precautions The U S Solid ...

Page 4: ...ble Handle the scale carefully It is a precision device subjecting it to impact may result in a malfunction If the device is not be used for a long period of time the power should be turned off and disconnect the power cable Do not disassemble remodel or repair this product or accessories Do not weigh items larger than the scale s range as this may damage the load cell of the scale Although the US...

Page 5: ...3 Design and Function 2 1 Components 1 Weighing Pan 2 Display Panel 3 Leveling Indicator 4 Keyboard 5 Leveling Feet 6 Rear Display Panel ...

Page 6: ...iption Switch on off the scale Enter setting while powering on Switch the setting item Perform calibration procedure with ZERO Zero the scale Perform calibration procedure with MODE Tare the scale Switch option in setting item ...

Page 7: ...e return to zero 5 Checkweighing mode the weight of the sample to be checked is lower than the lower limit 6 Checkweighing mode the weight of the sample to be checked is within the upper and lower limits 7 Checkweighing mode the weight of the sample to be checked is higher than the upper limit 8 Current weighing unit is lb 9 Current weighing unit is kg 10 Current weighing unit is g 11 Leveling ind...

Page 8: ...llowing accessories are part of the standard equipment for your new USS DBS93 94 Waterproof Bench Scale 1 Scale Main Body 1 Weighing Pan rectangular 1 Waterproof Power Cable 1 Instruction Manual 1 Quality Certification Check the instrument for damage in transit Immediately inform the U S Solid customer service if you have any complaints or parts are missing ...

Page 9: ... 1 Leveling the scale Accurate horizontal positioning is necessary for repeatable measurements and exact results To compensate for small irregularities or tilts at this location the instrument needs to be leveled Adjust the leveling feet of the bench scale until the air bubble in the indicator is centered The level indicator is located under the cover towards the rear of the bench scale ...

Page 10: ...of air bubble Adjustment method up Turn both feet clockwise down Turn both feet counterclockwise Left Turn left foot counterclockwise right foot clockwise Right Turn left foot clockwise right foot counterclockwise ...

Page 11: ...ing moving in the same room Changes in ambient conditions 1 Press TARE to zero the scale in the weighing state 2 Place the calibration weight on the weighing pan 3 Press MODE and ZERO simultaneously 4 Then CAL will be displayed 5 After the weight value of the calibration weight is displayed remove the calibration weight 6 The calibration procedure is completed and it will return to the weighing in...

Page 12: ...ing container on the weighing pan 2 Wait for the display to stabilize Then press TARE to tare and the displayed value will return to zero 3 Place the sample s to be weighed in the weighing container 4 After the display stabilizes read the value displayed Overweight Alarm When is displayed in the weighing state it indicates that the sample weight exceeds the maximum range of the scale The sample sh...

Page 13: ... to switch on the scale meanwhile long press MODE then F 1 will be displayed 2 Press MODE eight times continuously until P OFF is displayed 3 Then press TARE to switch to P ON 4 Press MODE two times continuously until END displays and returns to the weighing interface 5 Press POWER to switch off the scale 6 After several seconds press POWER to switch on again meanwhile long press MODE then F 1 wil...

Page 14: ... Press MODE and END will be displayed the lower limit and upper limit of checkweighing are set 13 After that the checkweighing operation can be performed When the sample weight is greater than the upper limit the Up light is on when the sample weight is less than the lower limit the Down light is on and when the sample weight is within the upper and lower limits range the Pass light is on ...

Page 15: ...weighing state press the UNIT button to switch weight units 14 In the power off state press POWER to switch on the scale while long press MODE then F 1 will be displayed 15 Then press MODE four times UN 00 will be displayed 16 Press TARE to switch options among UN 00 UN 01 and UN 02 00 01 and 02 stand for g kg and lb respectively 17 Press POWER to switch off the scale and the setting of weighing u...

Page 16: ...nt brightness 1 In the power off state press POWER to switch on the scale while long press MODE then F 1 will be displayed 2 Then press MODE five times br 0 will be displayed 3 Press TARE to switch options among br 0 br 1 br 2 and br 3 0 means the brightest 3 means the least bright 4 Press POWER to switch off the scale and the setting of screen brightness is completed ...

Page 17: ...abrasive cleaning agents etc or organic solvents to clean the scale Clean the scale using a piece of lint free cloth which has been wet with a mild detergent soap only 7 2 Disposal Disposal of equipment and packaging must be carried out by the operator in accordance with relevant laws of the country or region in which the equipment is to be used ...

Page 18: ...ower transformer damaged Instrument faulty Problem Weight display is constantly changing or unstable Possible causes Test bench vibration Electromagnetic field interference Power supply instability Problem Displayed value and actual weight are not consistent Possible causes Not returned to zero before weighing Scale not leveled well ...

Page 19: ... kg 15 kg Readability 5 g 1 g Repeatability 10 g 2 g Linearity Deviation 15 g 3 g Stabilization Time Within 3 seconds Pan Size 9 7 5 inches 235 190 mm Interface RS232 Power 110V 60Hz AC Dimensions 13 6 12 8 7 7 inches Net Weight 7 72 lbs Shipping Weight 8 82 lbs ...

Page 20: ...Contact Feel free to visit our website www ussolid com You can email us at service ussolid com You can call one of our friendly customer service representative at 1 800 209 4177 ...