background image

FV801 Rev.  A  06/21/21 

Page 9

   

 

For further assistance with troubleshooting visit http://www.zurn.com

Problem

Possible Cause

Corrective Action

Faucet is non-responsive Battery voltage low or no power.

Change out batteries or verify power provided by ACA

Sensor and electronic connector 

cable has bent or broken pins.

Check to see if pins are missing, bent, or broken. If yes, then the 

electronic box needs to be replaced

Sensor lens could have surface 

damage, scratches, or deposit 

build ups.

Inspect the sensor lens underneath the spout tip. May require

 

cleaning with a damp cloth or sensor replacement

Sensor lens could be obstructed 

by environmental factors such 

as highly reflective surfaces or 

bright lighting conditions.

If this is the case, then adjust the dip switch settings for sensor 

sensitivity per the diagram on previous page (switches 7 and 8). 

Direct sunlight into sink may make the unit inoperable. Do not 

allow direct sunlight into sink. 

 

 

If one mode does not help try one of the following:

 

-switching 7 to Off to set the sensor to the lowest sensitivity level 

(detection range may be less for all hand types).

 

-switching to alternate mode using switch 8; mode A or B depend-

ing on current configuration

 

 

Reset sensor after dip switch changes.

Faucet does not activate 

after initial installation

Sensor may not have had enough 

time to calibrate prior to a user 

or target trying to initiate activa-

tion.

Disconnect the sensor and electronic box cables and wait for 30 

seconds. Reconnect the cables, but allow for another 30 seconds 

before placing hands or a target within the sensor range. You 

should hear a click of the solenoid once calibration is complete.

Check if protective film was 

removed from lens.

Water does not stop

Electronics box cable connector 

has bent pins that are shorted

Check to see if pins are missing, bent, or broken. If yes, then the 

electronic box needs to be replaced.

Ghost Activation

Sensor lens could be obstructed 

or environmental factors such as 

highly reflective surfaces and/or 

bright sunlight conditions.

Check/clean sensor lens of any debris or buildup If sunlight or IR 

interference is a factor, refer to page 7 “Sensor Sensitivity” and 

adjust faucet for Mode B operation.  Also, decreasing the sensitiv-

ity setting (switch 7) may help.

 

 

Reset sensor after dip switch changes.

Sensor faucet is leaking

 

underneath the sink or

 

counter

Mounting hardware is not tight-

ened sufficiently (faucet shank, 

electronic box, supply hoses, etc)

Check and ensure the nuts and joints are tight and secure from 

the supply stop all the way to the faucet shank.

Missing the washer located in the 

solenoid box nut where it assem-

bles to the faucet shank.

Make sure the water supply stops are shut off. Check to ensure 

the washer is installed or damaged. If not, install washer. If a 

washer is damaged, replace with new.

Trouble Shooting:  Faucet

Summary of Contents for AquaSense Camaya Z6953-XL-W2

Page 1: ... of the product for his intended use and the purchaser assumes all risk and liability whatever in connection therewith Where permitted by law the implied warranty of merchantability is expressly excluded If the products sold hereunder are consumer products the implied warranty of merchantability is limited to a period of three years and shall be limited solely to the replacement of the defective g...

Page 2: ...ce to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving ante...

Page 3: ...nd finish Operating Water Pressure 10 125 psi Operational Water Temperature 33 F to 140 F 1 C to 60 C Aerator 0 5 GPM Vandal Resistant Standard Important Safety Information Installer is responsible for ensuring the product is installed and conforms to all plumbing codes and ordinances Do not convert or modify this Zurn product yourself All warranties will be voided Water supply lines must be sized...

Page 4: ... Gasket Aerator and Key Electronics Box with Solenoid Filter Spout Assembly Required Tools Batteries Batteries AquaSense Package Contents Adjustable Wrench Optional Smooth Jawed Wrench Phillips Head Screwdriver 3 32 M3 Allen Wrenches Filter Shank Mounting Kit Base Gasket Electronics Box with Solenoid Valve Hydrogenerator Y Cable Y Cable ...

Page 5: ...and or internal control and mixing valve 1 Carefully screw aerator into spout using included tool Take care to not pinch blue o ring 3 Ensure that slot in plastic washer is facing forward Sensor cable should pass through the front of the plastic mounting washer Spout Installation Instructions Z6953 XL W2 Z6956 XL W2 Cover Plate Installation Instructions 1 For faucets with CP4 or CP8 suffix Pass se...

Page 6: ...lied batteries Close lid and tighten lid screw 4 Connect sensor inline connector to electronics module and hydrogenerator Do not connect cable to the monitor at this step Use AA Lithium Batteries or AAAlkaline batteries but do not mix Lithium with Alkaline Monitor Electronic Module Installation Instructions 3 Connect the hydrogenerator A downstream of electronics module B and filter C to bottom of...

Page 7: ...mildly warm soapy water and then wipe the device dry with a clean soft towel or cloth Upon cleaning other areas of the restroom be sure the sensor lenses are protected from other cleaning chemicals solvents to prevent potential damages to the sensor and or electronics Care and Cleaning Instructions 6 Tighten electronics module assembly to shank and connect water line s 5 Open battery module using ...

Page 8: ...ctions Thick walls and reflective surfaces such as metal weaken the signal between the gateway and other devices The LoRa antenna is an omni directional antenna but for best results mount the device so the LoRa antenna is in a vertical position near the top of a wall We recommend conducting a site survey to test the signal strength in different locations before you mount the device Step 4 Mounting...

Page 9: ...sitivity is configured during pack out but the mode can be changed upon installation to optimize the performance within the sink Sensor sensitivity and sensor detection Modes A B may be tried for various lighting conditions sink materials and shape Switch 7 controls the sensitivity On default is normal sensitivity for most sinks providing max range Off set the sensor to a lower sensitivity level f...

Page 10: ... enough time to calibrate prior to a user or target trying to initiate activa tion Disconnect the sensor and electronic box cables and wait for 30 seconds Reconnect the cables but allow for another 30 seconds before placing hands or a target within the sensor range You should hear a click of the solenoid once calibration is complete Check if protective film was removed from lens Water does not sto...

Page 11: ...R sensor and Faucet actuator electronics Restore proper electrical connection as needed Step 2 Test activate the faucet If problem persists see Faucet Troubleshooting section in this guide for other steps Status Good posted in plumbSMART and faucet actuates but usage event data not appearing in plumbSMART Charts Step 1 Verify gateway is powered and LoRa Status LED is ON If not apply power per Zurn...

Page 12: ...acility IT administrator to modify port rules to allow Zurn gateway to access the internet via local router If gateway model is ETH LTE visually verify SIM card is installed in designated slot on the side of gateway If unit has a SIM card then disconnect ethernet cable and power cycle the gateway and wait for unit to initialize and connect LORA LED should turn ON solid after initialization Step 4 ...

Page 13: ...ect fixture model Verify product image and model number from plumbSMART product page Details tab against model number cited on product literature and packaging If incorrect contact Zurn Custom Service at 1 855 ONE ZURN 1 855 663 9876 for assistance High Daily Faucet Activate Rate posted in plumbSMART Step 1 Verify faucet is not falsely triggering See Faucet Troubleshooting section for appropriate ...

Page 14: ...To service solenoid use Phillips head screw driver to remove the screws on both sides of the electronics box With screws removed the electronics box can slide off and allow access to the solenoid for service Accessing the Solenoid ...

Page 15: ...pplied Adapter 3 8 Copper Tube By Others For use with MV Option Mixing Valve Filter and Mixing Valve Installation Control Box Attach adapter or Braided Hose SSH here with MV and TMV 1 options SSH suffix option P6900 12BC Filter Need to hold at pointed area by suitable wrench or by hand while applying torque to the supply hose nut Do not over torque the nut ...

Page 16: ...it Item 3 P6900 43 Hydrogenerator Assembly Item 4 P6900 HYD Electronics Box with Solenoid Item 5 P6950 XL B L Sensor Inline Connector Item 6 P6950 XL DC Description Product No Smart Connected Faucet Retrofit Kit Item 7 P6900 RK W2 8 Cast Cover Plate w Gasket Item 8 P6900 CP8 4 Cast Cover Plate w Gasket Item 9 P6900 CP4 Single Stainless Supply Hose XL Item 10 P6900 SH XL Stainless Supply Hoses XL I...

Reviews: