12
500°C
DC 9.2
weiter
500°C
DC 9.2
Abbruch
Dentin 1
Wash
DC 9.2_Wash_2022-05-03_14-19
2.1.5 Functional error during
program execution
Q
In the event of a functional error, the
oven interrupts the work process and
displays a window with an error
message in plain text and an error
code.
Q
This error message and the program
termination must be confirmed with
the green
OK
button.
Q
The error message that was
displayed last can be called up
later under
Settings
,
Last error
messages
.
2.1.6 End program
Q
The end of a program is indicated
by an acoustic signal and visually
by a text window. The name of the
completed program, the time and the
date at the end of the process are
given in the text window (image 12.2).
Q
A code is generated (image 12.3) for
documentation or quality assurance.
This consists of the serial number
of the furnace, the date YYMMDD -
time (24-hour system). The complete
program data can be called up under
Settings - Data transfer - Export
protocol files
(image 12.2). The last
50 program sequences are saved
here. Older program sequences are
overwritten. The log data should
therefore be backed up at regular
intervals.
Display, Status: Error message during a running program
Display, Status: Accomplished firing program
OK
OK
V-01
Abort: Vacuum error
Error
Program accomplished
Image 12.1
Image 12.3
Image 12.2
Run_Err: 0x00000100