Q00110 Rev A 8/06
Page 156 of 225
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
1st
2nd
3rd
5.
In a single-firm contact motion, touch (and hold) the stylus to the center of the 1
st
target
point until the
Z.ONE
detects this contact, and moves the calibration target point graphic to
the next (2
nd
) calibration position.
6.
Repeat the same single-firm touch (and hold) procedure for the remaining two calibration
points.
7.
If successful, the
Z.ONE
should return to the System Setup menu. If calibration failed
message appears, repeat the process being very attentive to the method of (single point) stylus
contact with the LCD display for contact.
18.5.2 Verifying Calibration
1.
Using the Menu Control, advance to the “Annotations” selection.
2.
Press the “SELECT”
button to bring up the Annotations screen.
3.
Press the “SELECT”
button to bring up the VKB (Virtual “Keyboard”).
4.
Using the trackball, position the arrow cursor in the “New Item” entry box on the screen and
press the “SET”
key.
5.
Using the stylus to touch within the virtual keys on the VKB, verify that the desired text
characters appear in the text entry box on the screen.