background image

ZOLL Base PowerCharger 

4x4

 Operator’s Manual

10

illuminate for the duration of the test cycle.

4. If the battery passes the test, the 

BATT. READY

 light 

will illuminate while the 

TESTING

 light remains lit. 

The battery is now recharged, ready for use, and 
capable of powering an M Series equipped with 
Pace/Defibrillation and Pulse Oximetry (SpO

2

) for 

approximately 1.5 hours in Monitor mode.

5. If a battery fails the test, the 

FAULT

 light will 

illuminate while the 

TESTING

 light remains lit. 

To identify the defect, press and hold the 

TEST

 

button while the battery is still in the compartment. 
Illumination of certain indicator lights together 
represent a specific defect; see “Fault Codes and 
Definitions” on page 11 and
 take corrective action. 
Repeat test.
Before disposing of the battery, test the battery 
compartment for possible defective operation (see 
“Battery Compartment Testing Procedure” on 
page 10). 

6. Record the date and results of the test on the battery 

label.

1RWH

Interrupting a charge cycle in progress by 
removing the battery will cause an audible 
warning of five (5) consecutive beeps to indicate 
that the battery is not fully charged and should 
not be used. All four indicator lights will illuminate 
for four (4) seconds and then go off. 

%DWWHU\&RPSDUWPHQW7HVWLQJ

3URFHGXUH

The ZOLL Base PowerCharger

4x4

 has four battery 

compartments that are able to test and charge batteries 
simultaneously and independently of each other. In 
addition, the ZOLL Base PowerCharger

4x4

 can test each 

battery compartment for defective operation through the 
simple press of a button.
To test a battery compartment:
1. Check to see that the unit is connected to live AC 

mains. The 

POWER

 light should be lit.

2. Remove the battery from the compartment to be 

tested.

3. Press the 

TEST 

button under the battery 

compartment.

4. All four indicator lights will illuminate for 

approximately 4-5 seconds.

5. If the 

TESTING

 light and 

BATT. READY 

light remain 

lit for three (3) seconds while the other lights go off, 
the compartment is fully operational and has no 
defects.

6. If the 

FAULT

 light is lit, the compartment is defective 

in some way.
To identify the defect, press and hold the 

TEST

 

button. Illumination of certain indicator lights together 
represent a specific defect. See “Fault Codes and 
Definitions” on page 11
 to identify the fault and take 
corrective action.

1RWH

Battery compartments operate independently of 
each other and one faulty compartment may not 
affect any others. However, there are common 
failures that can affect all compartments.

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

Summary of Contents for Base Powercharger 4x4

Page 1: ...ZOLL Base PowerCharger4x4 Operator s Manual Base PowerCharger 4x4 TM CAUTIO N DO NOT BLOCK VENTS OR PLACE ANYTHING ON TOP OF THE VENTS HEAT DAMAGE MAY OCCUR Part Number 9650 0054 Rev G...

Page 2: ...North America Consult your purchasing agreement for terms and conditions associated with your warranty Outside of North America consult the ZOLL authorized representative In order to maintain this wa...

Page 3: ...UH DWWHU RPSDUWPHQW 7HVWLQJ 3URFHGXUH DXOW RGHV DQG HILQLWLRQV 4XLFN 5HIHUHQFH KDUW IRU 2 DVH 3RZHU KDUJHU 4x4 ZLWK 4XLFN KDUJH 6HFWLRQ 2SHUDWLRQ XWR7HVW 2SHUDWLQJ RQWUROV DQG QGLFDWRUV 6HWXS DWWHU KD...

Page 4: ...ZOLL Base PowerCharger 4x4 Operator s Manual ii This page intentionally left blank...

Page 5: ...er 4x4 with AutoTest automatically tests battery capacity with each battery recharge In addition the ZOLL Base PowerCharger 4x4 with AutoTest illuminates the Batt Ready indicator when the battery is f...

Page 6: ...HVKRRWLQJ This section describes several routine checks to assure proper unit operation It includes cleaning and maintenance instructions and troubleshooting guides intended for use by non technical p...

Page 7: ...This warranty does not cover items subject to normal wear and burnout during use including but not limited to lamps fuses batteries patient cables and accessories d The foregoing warranty constitutes...

Page 8: ...epeated use of partially charged batteries will reduce their capacity and operating life Do not use a battery unless the BATT READY indicator is lit and the FAULT light is off Failure to do so may res...

Page 9: ...compartment FAULT Illuminates when battery capacity test or battery compartment test fails Separate indicator light for each battery compartment TESTING Illuminates when either a battery test or a bat...

Page 10: ...order to allow tracking of loan equipment You will be given a Return Claim Number under which to return the ZOLL Base PowerCharger 4x4 Send the unit in its original packaging to ZOLL Medical Corporati...

Page 11: ...x4 refer to QuickCharge unless otherwise specified If your device does not indicate QuickCharge or AutoTest on the label in the upper right corner refer to the QuickCharge section of this manual for p...

Page 12: ...1 Illuminates when either a battery capacity test cycle is in progress or a battery compartment test is completed Illuminates with the BATT READY light indicating a usable fully charged battery capab...

Page 13: ...nsecutive beeps to indicate that the battery is not fully charged and should not be used In addition all four indicator lights will illuminate for four 4 seconds and then go off Repeated use of partia...

Page 14: ...tor lights will illuminate for four 4 seconds and then go off DWWHU RPSDUWPHQW 7HVWLQJ 3URFHGXUH The ZOLL Base PowerCharger 4x4 has four battery compartments that are able to test and charge batteries...

Page 15: ...d range Low Constant Current Amount of current during a charge was below the specified range High Constant Current Amount of current during a charge was above the specified range VCC Voltage Internal...

Page 16: ...e ZOLL Base PowerCharger 4x4 will beep to indicate the battery has been removed and should NOT be used The POWER light should be on as long as the unit is plugged into a live AC source 23 5 7 21 4XLFN...

Page 17: ...r 4x4 refer to AutoTest unless otherwise specified If your device does not indicate QuickCharge or AutoTest on the label in the upper right corner refer to the QuickCharge section of this manual for p...

Page 18: ...tment or any other failure occurs There are four FAULT indicator lights Each one corresponds to the battery charging testing compartment directly above 7 67 1 Illuminates when either a battery capacit...

Page 19: ...cific defect see Fault Codes and Definitions on page 17 and take corrective action Before disposing of the battery test the battery compartment for defective operation see Battery Compartment Testing...

Page 20: ...ry will cause an audible warning of five 5 consecutive beeps to indicate that the battery is not fully charged and should not be used All four indicator lights will illuminate for four 4 seconds and t...

Page 21: ...range Low Constant Current Amount of current during a charge was below the specified range High Constant Current Amount of current during a charge was above the specified range VCC Voltage Internal fi...

Page 22: ...ger 4x4 will beep to indicate the battery has been removed and should NOT be used The POWER light should be on as long as the unit is plugged into a live AC source 23 5 7 21 XWR7HVW 5 5 21 77 5 8 7 7...

Page 23: ...all battery compartments empty check that all indicator lights except POWER are off 3 Press the TEST button under each compartment to verify that the compartments are fully operational see Battery Com...

Page 24: ...r power source Check for blown fuse 2 CHARGER ON indicator light does not illuminate when battery is in compartment Check that battery is properly seated in compartment Check that battery contacts are...

Page 25: ...ent refer to the ZOLL Battery Management Program part number 9650 0019 2 DWWHU 3DFNV ZOLL PD 4410 Smart Smart Ready Battery and XL Battery Packs are five cell assemblies of sealed lead acid batteries...

Page 26: ...ZOLL Base PowerCharger 4x4 Operator s Manual 22 This page intentionally left blank...

Page 27: ......

Page 28: ...336 ZOLL Medical France 189 Rue Nationale 10 78310 Coignieres FRANCE Telephone 33 1 30 05 14 95 Telefax 33 1 30 49 06 93 Middle East Africa Regional Office 14 Persefonis Ekali 14578 Athens Greece Tele...

Reviews: