ZOLL Base PowerCharger
4x4
Operator’s Manual
10
illuminate for the duration of the test cycle.
4. If the battery passes the test, the
BATT. READY
light
will illuminate while the
TESTING
light remains lit.
The battery is now recharged, ready for use, and
capable of powering an M Series equipped with
Pace/Defibrillation and Pulse Oximetry (SpO
2
) for
approximately 1.5 hours in Monitor mode.
5. If a battery fails the test, the
FAULT
light will
illuminate while the
TESTING
light remains lit.
To identify the defect, press and hold the
TEST
button while the battery is still in the compartment.
Illumination of certain indicator lights together
represent a specific defect; see “Fault Codes and
Definitions” on page 11 and take corrective action.
Repeat test.
Before disposing of the battery, test the battery
compartment for possible defective operation (see
“Battery Compartment Testing Procedure” on
page 10).
6. Record the date and results of the test on the battery
label.
1RWH
Interrupting a charge cycle in progress by
removing the battery will cause an audible
warning of five (5) consecutive beeps to indicate
that the battery is not fully charged and should
not be used. All four indicator lights will illuminate
for four (4) seconds and then go off.
%DWWHU\&RPSDUWPHQW7HVWLQJ
3URFHGXUH
The ZOLL Base PowerCharger
4x4
has four battery
compartments that are able to test and charge batteries
simultaneously and independently of each other. In
addition, the ZOLL Base PowerCharger
4x4
can test each
battery compartment for defective operation through the
simple press of a button.
To test a battery compartment:
1. Check to see that the unit is connected to live AC
mains. The
POWER
light should be lit.
2. Remove the battery from the compartment to be
tested.
3. Press the
TEST
button under the battery
compartment.
4. All four indicator lights will illuminate for
approximately 4-5 seconds.
5. If the
TESTING
light and
BATT. READY
light remain
lit for three (3) seconds while the other lights go off,
the compartment is fully operational and has no
defects.
6. If the
FAULT
light is lit, the compartment is defective
in some way.
To identify the defect, press and hold the
TEST
button. Illumination of certain indicator lights together
represent a specific defect. See “Fault Codes and
Definitions” on page 11 to identify the fault and take
corrective action.
1RWH
Battery compartments operate independently of
each other and one faulty compartment may not
affect any others. However, there are common
failures that can affect all compartments.
TEST
TESTING
FAULT
BATT. READY
CHARGER ON