background image

9

Operation (QuickCharge)

To turn the ZOLL Base PowerCharger

4x4

 on, plug the 

power cord into live AC mains. All indicator lights will 
illuminate for approximately five (5)  seconds and then 
go off. An audible tone will sound indicating self test 
passed and the 

POWER

 light will remain lit.

To charge a battery:
1. Check to see that the unit is connected to live AC 

mains. The 

POWER

 light should be lit.

2. Insert a battery into any compartment. Make sure the 

battery is fully seated (a click can be heard).

3. The 

CHARGER ON 

indicator illuminates and remains 

on during the charging cycle.

4. In four (4) hours or less for the PD4410/Smart/

Smart Ready Battery and eight (8) hours or less for 
all XL Batteries, the 

CHARGER ON 

light will go off 

and the 

BATT. READY 

light will illuminate indicating 

the battery has been charged to 100% of its present 
capacity and is ready for use.

If the battery or battery compartment is not capable of 
charging, the 

FAULT

 light will illuminate and the 

CHARGER ON

 light may stay on.

To identify the defect, press and hold the 

TEST

 

button while the battery is still in the compartment. 

Illumination of certain indicator lights together 
represent a specific defect; see “Fault Codes and 
Definitions” on page 11 and
 take corrective action. 
Before disposing of the battery, test the battery 
compartment for defective operation (see “Battery 
Compartment Testing Procedure” on page 10) and 
perform a battery test. 

1RWH

Interrupting a charge cycle in progress by 
removing the battery will cause an audible 
warning of five (5) consecutive beeps to indicate 
that the battery is not fully charged and should not 
be used. In addition, all four indicator lights will 
illuminate for four (4) seconds and then go off. 

Repeated use of partially charged batteries will 
reduce battery capacity and operating life.

5. To remove the battery, insert a finger into the recess 

at the top of the battery compartment, press against 
the battery to disengage the battery locking clip and 
lift the battery out. A compartment test is performed 
illuminating all indicator lights, briefly followed by the 

BATT. READY

 and 

TESTING

 indicator lights 

illuminating for two (2) seconds and then all indicator 
lights going off, except the 

POWER

 indicator light.

0DQXDO%DWWHU\7HVWLQJ3URFHGXUH

ZOLL batteries are sealed lead-acid batteries. It is 
recommended that all batteries be tested for capacity at 
least every 90 days and that the test results and date be 
written on the battery label. 
All four battery compartments can test batteries. It is not 
necessary to insert a fully charged battery for testing; the 
ZOLL Base PowerCharger

4x4

 will automatically charge 

the battery first, test it, and then recharge the battery to 
full capacity. Testing a battery in one compartment will 
not affect any other charge or testing cycles in progress.
To test a battery:

1. Check to see that the ZOLL Base PowerCharger

4x4

 

is connected to live AC mains. The 

POWER

 indicator 

light should be lit.

2. Insert a battery into any compartment. Make sure the 

battery is fully seated (a click can be heard).

3. Press the 

TEST

 button for the appropriate 

compartment. The 

TESTING

 indicator light will 

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

Summary of Contents for Base Powercharger 4x4

Page 1: ...ZOLL Base PowerCharger4x4 Operator s Manual Base PowerCharger 4x4 TM CAUTIO N DO NOT BLOCK VENTS OR PLACE ANYTHING ON TOP OF THE VENTS HEAT DAMAGE MAY OCCUR Part Number 9650 0054 Rev G...

Page 2: ...North America Consult your purchasing agreement for terms and conditions associated with your warranty Outside of North America consult the ZOLL authorized representative In order to maintain this wa...

Page 3: ...UH DWWHU RPSDUWPHQW 7HVWLQJ 3URFHGXUH DXOW RGHV DQG HILQLWLRQV 4XLFN 5HIHUHQFH KDUW IRU 2 DVH 3RZHU KDUJHU 4x4 ZLWK 4XLFN KDUJH 6HFWLRQ 2SHUDWLRQ XWR7HVW 2SHUDWLQJ RQWUROV DQG QGLFDWRUV 6HWXS DWWHU KD...

Page 4: ...ZOLL Base PowerCharger 4x4 Operator s Manual ii This page intentionally left blank...

Page 5: ...er 4x4 with AutoTest automatically tests battery capacity with each battery recharge In addition the ZOLL Base PowerCharger 4x4 with AutoTest illuminates the Batt Ready indicator when the battery is f...

Page 6: ...HVKRRWLQJ This section describes several routine checks to assure proper unit operation It includes cleaning and maintenance instructions and troubleshooting guides intended for use by non technical p...

Page 7: ...This warranty does not cover items subject to normal wear and burnout during use including but not limited to lamps fuses batteries patient cables and accessories d The foregoing warranty constitutes...

Page 8: ...epeated use of partially charged batteries will reduce their capacity and operating life Do not use a battery unless the BATT READY indicator is lit and the FAULT light is off Failure to do so may res...

Page 9: ...compartment FAULT Illuminates when battery capacity test or battery compartment test fails Separate indicator light for each battery compartment TESTING Illuminates when either a battery test or a bat...

Page 10: ...order to allow tracking of loan equipment You will be given a Return Claim Number under which to return the ZOLL Base PowerCharger 4x4 Send the unit in its original packaging to ZOLL Medical Corporati...

Page 11: ...x4 refer to QuickCharge unless otherwise specified If your device does not indicate QuickCharge or AutoTest on the label in the upper right corner refer to the QuickCharge section of this manual for p...

Page 12: ...1 Illuminates when either a battery capacity test cycle is in progress or a battery compartment test is completed Illuminates with the BATT READY light indicating a usable fully charged battery capab...

Page 13: ...nsecutive beeps to indicate that the battery is not fully charged and should not be used In addition all four indicator lights will illuminate for four 4 seconds and then go off Repeated use of partia...

Page 14: ...tor lights will illuminate for four 4 seconds and then go off DWWHU RPSDUWPHQW 7HVWLQJ 3URFHGXUH The ZOLL Base PowerCharger 4x4 has four battery compartments that are able to test and charge batteries...

Page 15: ...d range Low Constant Current Amount of current during a charge was below the specified range High Constant Current Amount of current during a charge was above the specified range VCC Voltage Internal...

Page 16: ...e ZOLL Base PowerCharger 4x4 will beep to indicate the battery has been removed and should NOT be used The POWER light should be on as long as the unit is plugged into a live AC source 23 5 7 21 4XLFN...

Page 17: ...r 4x4 refer to AutoTest unless otherwise specified If your device does not indicate QuickCharge or AutoTest on the label in the upper right corner refer to the QuickCharge section of this manual for p...

Page 18: ...tment or any other failure occurs There are four FAULT indicator lights Each one corresponds to the battery charging testing compartment directly above 7 67 1 Illuminates when either a battery capacit...

Page 19: ...cific defect see Fault Codes and Definitions on page 17 and take corrective action Before disposing of the battery test the battery compartment for defective operation see Battery Compartment Testing...

Page 20: ...ry will cause an audible warning of five 5 consecutive beeps to indicate that the battery is not fully charged and should not be used All four indicator lights will illuminate for four 4 seconds and t...

Page 21: ...range Low Constant Current Amount of current during a charge was below the specified range High Constant Current Amount of current during a charge was above the specified range VCC Voltage Internal fi...

Page 22: ...ger 4x4 will beep to indicate the battery has been removed and should NOT be used The POWER light should be on as long as the unit is plugged into a live AC source 23 5 7 21 XWR7HVW 5 5 21 77 5 8 7 7...

Page 23: ...all battery compartments empty check that all indicator lights except POWER are off 3 Press the TEST button under each compartment to verify that the compartments are fully operational see Battery Com...

Page 24: ...r power source Check for blown fuse 2 CHARGER ON indicator light does not illuminate when battery is in compartment Check that battery is properly seated in compartment Check that battery contacts are...

Page 25: ...ent refer to the ZOLL Battery Management Program part number 9650 0019 2 DWWHU 3DFNV ZOLL PD 4410 Smart Smart Ready Battery and XL Battery Packs are five cell assemblies of sealed lead acid batteries...

Page 26: ...ZOLL Base PowerCharger 4x4 Operator s Manual 22 This page intentionally left blank...

Page 27: ......

Page 28: ...336 ZOLL Medical France 189 Rue Nationale 10 78310 Coignieres FRANCE Telephone 33 1 30 05 14 95 Telefax 33 1 30 49 06 93 Middle East Africa Regional Office 14 Persefonis Ekali 14578 Athens Greece Tele...

Reviews: