9
Operation (QuickCharge)
To turn the ZOLL Base PowerCharger
4x4
on, plug the
power cord into live AC mains. All indicator lights will
illuminate for approximately five (5) seconds and then
go off. An audible tone will sound indicating self test
passed and the
POWER
light will remain lit.
To charge a battery:
1. Check to see that the unit is connected to live AC
mains. The
POWER
light should be lit.
2. Insert a battery into any compartment. Make sure the
battery is fully seated (a click can be heard).
3. The
CHARGER ON
indicator illuminates and remains
on during the charging cycle.
4. In four (4) hours or less for the PD4410/Smart/
Smart Ready Battery and eight (8) hours or less for
all XL Batteries, the
CHARGER ON
light will go off
and the
BATT. READY
light will illuminate indicating
the battery has been charged to 100% of its present
capacity and is ready for use.
If the battery or battery compartment is not capable of
charging, the
FAULT
light will illuminate and the
CHARGER ON
light may stay on.
To identify the defect, press and hold the
TEST
button while the battery is still in the compartment.
Illumination of certain indicator lights together
represent a specific defect; see “Fault Codes and
Definitions” on page 11 and take corrective action.
Before disposing of the battery, test the battery
compartment for defective operation (see “Battery
Compartment Testing Procedure” on page 10) and
perform a battery test.
1RWH
Interrupting a charge cycle in progress by
removing the battery will cause an audible
warning of five (5) consecutive beeps to indicate
that the battery is not fully charged and should not
be used. In addition, all four indicator lights will
illuminate for four (4) seconds and then go off.
Repeated use of partially charged batteries will
reduce battery capacity and operating life.
5. To remove the battery, insert a finger into the recess
at the top of the battery compartment, press against
the battery to disengage the battery locking clip and
lift the battery out. A compartment test is performed
illuminating all indicator lights, briefly followed by the
BATT. READY
and
TESTING
indicator lights
illuminating for two (2) seconds and then all indicator
lights going off, except the
POWER
indicator light.
0DQXDO%DWWHU\7HVWLQJ3URFHGXUH
ZOLL batteries are sealed lead-acid batteries. It is
recommended that all batteries be tested for capacity at
least every 90 days and that the test results and date be
written on the battery label.
All four battery compartments can test batteries. It is not
necessary to insert a fully charged battery for testing; the
ZOLL Base PowerCharger
4x4
will automatically charge
the battery first, test it, and then recharge the battery to
full capacity. Testing a battery in one compartment will
not affect any other charge or testing cycles in progress.
To test a battery:
1. Check to see that the ZOLL Base PowerCharger
4x4
is connected to live AC mains. The
POWER
indicator
light should be lit.
2. Insert a battery into any compartment. Make sure the
battery is fully seated (a click can be heard).
3. Press the
TEST
button for the appropriate
compartment. The
TESTING
indicator light will
TEST
TESTING
FAULT
BATT. READY
CHARGER ON
TEST
TESTING
FAULT
BATT. READY
CHARGER ON
TEST
TESTING
FAULT
BATT. READY
CHARGER ON