13
Section 3
Maintenance and Troubleshooting
Following is a list of nontechnical operational checks. These
checks should be performed periodically to ensure proper
equipment operation. If during this procedure, the operator
discovers a malfunction, contact ZOLL Technical Service.
,QVSHFWLRQ
Assure that the unit is clean (with no fluid spills) and nothing
is stored on the unit.
Inspect all batteries for good condition.
&OHDQLQJ
Clean the Base PowerCharger
1x1
, with a soft cloth, mild
soap and water.
Do not immerse any part of the Base PowerCharger
1x1
in
water. Do not use alcohol or ketones (MEK, acetone, etc.).
Do not autoclave the Base PowerCharger
1x1
.
2SHUDWLRQDO&KHFNV
1. Connect the unit to live AC mains. All four indicator lights
will illuminate for approximately 4-5 seconds.The
POWER
indicator will remain lit while the other lights go
off.
2. With the battery compartment empty, check that all
indicator lights except
POWER
are off.
3. Press the
TEST
button to check that the unit is fully
operational.
4. All four indicator lights will illuminate for approximately 4-
5 seconds.
5. If the
TESTING
light and
BATT. READY
light remain lit
for three seconds while the other lights go off, the unit is
fully operational without any defects.
6. If the
FAULT
light remains lit, the unit is defective.
,QVHUWDEDWWHU\LQWRWKHFRPSDUWPHQW9HULI\WKDWWKH
&+$5*(521
OLJKWLOOXPLQDWHVLPPHGLDWHO\
7URXEOHVKRRWLQJ
The troubleshooting guide provided below is intended for
use by nontechnical personnel during operation. This
section answers many of the common problems or
questions that arise during operation.
If trouble persists after consulting this guide, call ZOLL and
ask for the Technical Service Department.
1.
POWER
light does not illuminate.
•
Check that unit is plugged into an appropriate power
source.
•
Check for defective power cord.
•
Plug unit into another power source.
•
Check for blown fuse.
2.
CHARGER ON
indicator light does not illuminate when
battery is in compartment.
•
Check that battery is properly seated in compartment.
•
Check that battery contacts are clean and not
damaged.
•
Check that unit is plugged into an appropriate power
source.
•
Test the battery compartment for functionality.
•
Place another battery in the compartment.
3.
TESTING
light does not illuminate when TEST button is
pressed.
•
Check that battery is properly seated in compartment.
•
Check that unit is plugged into an appropriate power
source.
•
Verify that battery compartment is fully operational with
no defects.
•
Repeat test cycle.
•
,IEDWWHU\IDLOVVHFRQGWHVWUHPRYHEDWWHU\IURP
use.
Summary of Contents for Base PowerCharger 1x1
Page 1: ...ZOLL Base PowerCharger1x1 Operator s Manual 3DUW 1XPEHU 5HY...
Page 10: ...ZOLL Base PowerCharger 1x1 Operator s Manual 6 This page intentionally left blank...
Page 16: ...ZOLL Base PowerCharger 1x1 Operator s Manual 12 This page intentionally left blank...
Page 18: ...ZOLL Base PowerCharger 1x1 Operator s Manual 14 This page intentionally left blank...
Page 20: ...ZOLL Base PowerCharger 1x1 Operator s Manual 16 This page intentionally left blank...
Page 25: ......