5.2.1
User Advisory (45)
Figure 5-5
User Advisory (45)
The Platform driveshaft has a “home” position that is a point of reference for Platform operation. If the
driveshaft is not at its home position when the Platform is powered on, a User Advisory (45) will occur.
This User Advisory will persist until the driveshaft is returned to its home position.
To clear a User Advisory (45) pull up on the LifeBand until the chest bands are fully extended (thereby
moving the driveshaft back to its home position), and then RESTART.
If a User Advisory (45) can't be cleared, Power off the Platform and contact ZOLL at +1.800.348.9011
or +1.978.421.9655.
Warning:
Removing the band clip when the driveshaft is not at its home position (see Figure 2-8 on
page 2-7) will result in a permanent User Advisory (45) that the user will not be able to
clear. To avoid this situation, the following guidelines should be adhered to:
1.
The LifeBand should be removed from the driveshaft ONLY from its home position.
2.
The LifeBand must be completely unwound with the seam visible (see Figure 2-10 on
page 2-8).
3.
If the LifeBand is cut, care should be taken to ensure that the bands are fully extended
before the cover plate is opened and the band clip is removed.
4.
Do not attempt to defeat the shaft lock mechanism, which keeps the driveshaft at its
home position when a LifeBand is not installed.
3 0
:
2
m e n u
RESTART
P u l l u p c h e s t b a n d ,
a n d p r e s s R E S T A R T
.
USER ADVISORY(45)
Summary of Contents for AutoPulse 100
Page 1: ...P N 12555 001 Rev 9 Resuscitation System Model 100 User Guide...
Page 8: ...Page viii User Guide This page left intentionally blank...
Page 38: ...Page 2 16 User Guide This page left intentionally blank...
Page 66: ...Page 4 6 User Guide This page left intentionally blank...
Page 72: ...Page 5 6 User Guide This page left intentionally blank...
Page 74: ...Page A 2 User Guide This page left intentionally blank...