25
3.5.2
Turning the system on
3.5.2.1
Acclimatisation period
After being transported, the unit requires one hour for each 2.5º of heating or cooling if the temperature was below
10ºC or above 40ºC.
NOTE:
Please keep at least 20 to 30 cm of wall clearance at the back of the system to ensure good ventilation. Oth-
erwise, malfunctions may occur as the temperature increases inside the unit.
3.5.2.2
Connecting to the power supply
After ensuring that the clinic/practice's power supply is normal and that the AC voltage type meets the power require-
ments indicated on the system label, plug the power cable's connector into the power socket on the back of the system
and the other end of the power cable into the AC power socket in the clinic/practice.
Please use the power cable provided by the manufacturer; other types of power cable are not permitted.
Press
for 1 second to start the system.
To shut down, press
and the pop-up shutdown dialogue pops up.
Press the Enter key, and the unit turns off.
Press
for 3 seconds and the system is forcibly shut down.
CAUTION:
If the system is connected using the wrong power supply unit, the incorrect power supply may cause dam-
age to the system and injuries to users.
3.5.3
Probe installation
Caution:
Please only use the probes provided by the manufacturer for this model; other types of probes are not permit-
ted for use with this system! Otherwise, this can lead to damage to the system and probe.
CAUTION:
Before connecting the probe, please carefully check the probe lens, probe cable and probe connector for
abnormal functionality or conditions, such as cracks or malfunctions. Probes that function abnormally or ap-
pear abnormal are not allowed to be connected to the system; otherwise there is the possibility of electricity
shock.
Hold the lock switch for the probe connector and insert the connection socket vertically. Release the probe
lock switch. Check the probe lock with one hand to make sure that it is not loose and is securely connected.
CAUTION:
The probe can only be installed/dismounted when the power supply is in the “switched off” state, other-
wise the machine or probe can get damaged. When installing and dismounting the probe, please place
the probe head inside the holder provided to prevent the probe getting damaged by falling down.