23
3.5.1.2
Environmental conditions for transport and storage
The following environmental conditions for transport and storage fall within the system tolerances:
Temperature: from 5° to 40°C
Relative humidity: ≤80% RH non-condensing
Atmospheric pressure: 700 hPa ~ 1060 hPa
3.5.1.3
Electrical requirements
Power consumption
< 60 VA
WARNING:
Maintain a fluctuation range of less than ±10% in accordance with the voltage indication on the
rear of the system, otherwise the system may get damaged.
NOTE:
Please observe the described power requirements. Only use power cables that meet the system guidelines – failure
to follow these procedures may lead to system damage.
The power supply may vary at different geographical locations. Refer to the detailed power ratings at the rear of the
system for detailed information.
Battery
To avoid the battery bursting or igniting, or fumes from the battery which can cause damage to the equipment,
please observe the following precautions: Do not immerse the battery in water or allow it to get wet. Do not place
the battery in a microwave oven or pressurised container. If the battery leaks or emits an odour, remove it from all
possible inflammable sources. If the battery emits an odour or heat, becomes deformed or discoloured, or appears
abnormal in any way during use, recharging or storage, immediately remove it and stop using it. If you still have
questions about the battery and short-term storage of the battery pack (less than a month): Store the battery in the
temperature range between 0°C (32°F) and 50°C (122°F).
Long-term storage of the battery pack (3 months or more): Store the battery in the temperature range between -
20ºC (-4ºF) and +45ºC (113ºF); upon receipt of the SONIDOSMART and before the first use, it is recommended that
the customer complete a full charge/discharge cycle. If the battery is not used for a period of more than two months,
the customer should also complete a full charge/discharge cycle. It is also recommended to store the battery in a
shaded and cool area at FCC (full current capacity). • One complete charge/discharge cycle procedure: 1. Full dis-
charge of the battery to automatically and fully power down the SONIDOSMART unit.2. Charge the SONIDOSMART
to 100% FCC (full current capacity). 3. Discharge of the Venue 40 for complete shutdown (takes one hour to dis-
charge).• When storing packs for more than 6 months, charge the pack at least once every six months to prevent
battery leakage and deterioration in performance.
3.5.1.4
Operating space
Please leave enough free space at the back of the system to ensure good ventilation.
.
CAUTION:
Leave enough space at the back of the system. Otherwise, malfunctions may occur as the temperature
increases inside the unit.