Cleaning
Disinfection
10
Page 24
the inner and outer surfaces of the protective are wetted. Leave the protective
cape in the solution as long as defined by the manufacturer of the disinfection
agent. Rinse off the protection cap under running water.
Cleaning / Disinfection, by machine:
Preparation:
Visible contaminations must be removed by manually before performing
cleaning / disinfection. Proceed as described above.
Performance:
Perform a cleaning and disinfection by machine using the following
parameters:
Cleaning agent: neodisher® MediClean forte (manufacturer: Dr. Weigert)
Cleaning: 10 minutes at 55°C
Disinfection: 5 minutes at 93°C
Caution:
If flammable solutions are used for cleaning and disinfecting,
sufficient time must be allowed for the solution to evaporation before using the
device. Otherwise, it may lead to inflammation.
Suitable Disinfection
agents
For manual disinfection, the following agents are suitable:
mikrozid® sensitiv wipes
Antifect FF
Gigasept FF
Quartamon Med
Observe the application information of the corresponding manufacturer.
Note:
Only use the device in a completely hygienic environment.
Caution
!
When using lubricants, the silicone protection cap must be placed over the
applicator head in order to avoid contamination.
If the device is used without the cap, lubricant may enter the interior of the
applicator head and the handpiece, which can lead to permanent
contamination and malfunction.
Note:
In this case, the warranty is null and void.