Operating Instructions
8.4 Display and Buttons
8
Page 14
Description of the
display elements
and buttons
(1) Pulse energy
Displays the pulse energy setting. During active therapy, the bar graph is
filled.
The pulse energy can be set both before and during the delivery of pulses. The
pulse energy is adjustable between 60 and 185 mJ, in 10mJ increments.
(2) Mode
Displays the selected mode. Upon activation of the window, the selection menu
with the operating modes is displayed:
Series Pulse, Burst 4 Pulse, Burst 8 Pulse, Burst 12 Pulse.
The desired mode of operation is selected directly on the corresponding line.
(3) Frequency
Displays the set frequency.
Frequency range: 1 Hz - 22 Hz, adjustable using the right controller. Here the
maximum selectable frequency depends on the selected power level.
(4) Saving
Pressing the key opens the field to enter the individual name of a program. The
program is automatically stored in the favourites list.
(5) Counting
direction
Pressing the button sets the counting direction (ascending or descending) for
the number of pulses emitted.
(6) Pulse count
Display of the preselected number of pulses and the currently emitted pulses
as well as the total number of pulses emitted in the case of non pre-set
pulse number.
Ascending or descending display the counting direction.
Pressing the pulse number field opens the entry menu to enter a
pre-set number of pulses. The pre-set is done in increments of 100.
(7) Back
For ascending counting direction, this resets to 0; for descending counting
direction, this resets to the pre-set number of pulses.
1
2
3
4
5
7
9
8
6
10