Manufacturer's EMC Declaration
17
Page 36
Key features of the en
Puls
Version 2.0 are: interference-free delivery of radial pulses, trouble-free operation of all functions.
Continuous operation is not required for the intended use.
Guidelines and manufacturer's declaration - electromagnetic immunity
The en
Puls
Version 2.0 device is intended for use in the electromagnetic environment specified below. The customer or the
user of the en
Puls
Version 2.0 device should ensure that it is used in such an environment.
Immunity
tests
IEC 60601 - Test
Level
Compliance
level
Electromagnetic environment - Guidelines
Conducted
RF
disturbance
variables in
accordance
with IEC
61000-4-6
Radiated
RF
disturbance
variables in
accordance
with IEC
61000-4-3
3 V
RMS
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
RMS
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Portable and mobile RF communications equipment
should not be used any closer to the en
Puls
Version
2.0
,
including cables, than the recommended
separation distance calculated using the equation
relevant for the transmission frequency.
Recommended separation distance:
d = 1.2 √P
d = 0.35 √P for 80 MHz to 800 MHz
d = 0.7 √P for 800 MHz to 2.5 GHz
where P is the power rating of the transmitter in
watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in
metres (m).
At all frequencies, the field strengths of fixed RF
transmitters should, according to an on-site
a
investigation, be less than the compliance level
b
.
Faults may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 Hz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all cases. The electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.