Safety / Maintenance / Service
12.2 Service
12
Page 23
Performing S01
Activating the “Start” button under the “Run S01” display field starts the “Run
S01” maintenance program.
Description of Run S01 (total time approx. 90 minutes):
This program simulates two 15-minute treatment sessions and three cooling
processes, as well as a defrosting process. The program records key
parameters such as temperature and duration. The current temperatures of
the compressor and evaporator can be checked during the test.
At the end, a summary appears on the screen. You should photograph or write
down the summary of the data at the end of Run S01.
Caution! Run S01 lasts about 90 minutes. Afterwards, Cryo 7 restarts.
Performing S02
Activa
ting the “Start” button under the “Run S02” display field starts the “Run
S02” maintenance program.
Description of Run S02:
The device simulates a 15-minute therapy session.
Run S02 is repeated until the start/stop button is pressed.
Run S02 is a stress test for the cooling system.
Please use Run S02 only with instructions from a technician. Afterwards, Cryo
7 performs a restart.
Software versions
Activating the “Software versions” button displays a list of the software
versions and serial numbers.
Operating hours
This display gives an overview of various operating hours, such as availability
if service is needed and the Cryo 7 configuration (derma or PT).