
Safety / Maintenance / Service
12.1 Safety
12
Page 21
Safety and
maintenance
Cryo 7 was developed and manufactured in compliance with the safety
regulations of IEC 60601-1.
Zimmer MedizinSysteme GmbH is responsible for the safety and reliability of
the device only if
1.
the device is operated using a proper power outlet with a protective
contact and the electrical installation complies with DIN VDE 0100 part
710.
2.
the device is operated in accordance with the instructions.
3.
expansions, readjustments or modifications are carried out only by
persons authorised by Zimmer MedizinSysteme.
4.
the user was instructed prior to use in the functional safety, the proper
operating condition, and the mechanical integrity of the device and
accessories.
5.
the device is operated only by properly trained personnel.
6.
the device is not operated in areas at risk of explosion and/or in a
flammable atmosphere.
7.
the device is immediately disconnected from the mains if liquid
penetrates it.
8.
the device does not contain any parts that can be serviced or repaired
by the operator. This does not apply to the defrost container (emptying
and cleaning) as well as to the replacement of the air filter. If the air
filter needs to be changed, this is indicated in the status bar.
9.
Replacement of the lithium battery by untrained personnel can lead to
hazards.
Reporting
All serious incidents associated with the product are to be reported to the
manufacturer and the competent authority of the state in which the user and/or
the patient is located.