Application information and warnings
2
Page 4
If the operator identifies a potential safety issue or operating abnormality during
use, the operator should stop the treatment immediately and contact the
manufacturer .
To avoid the risk of electric shock, the mains plug must be disconnected from
the power supply before performing cleaning or maintenance work.
Check Cryo 7 before use. If it has any damage, the device should not be used
and the customer service needs to be contacted.
Do not place the device in the vicinity of heat sources (heater, hot medicinal
mud products, sauna, etc.) and leave a distance of at least 50 cm between the
device and a wall (to enable the supply of cool air).
If used in combination with a laser device, the laser ventilation system should
not interfere the cooling of the cold air therapy device.
Do not switch Cryo 7 on and immediately off. Doing so could damage Cryo 7.
To transport over thresholds, Cryo 7 must be moved over a threshold
backwards by lifting it at the handle.
After transport or storage on its side, Cryo 7 must be brought to an upright
position and kept in an upright position for at least 30 minutes before it is
switched on. Otherwise the compressor will be damaged.
Before use, ensure that Cryo 7 is supplied with power via a properly grounded
power outlet (electrical installation according to DIN VDE 0100 part 710). Cryo
7 may only be operated with the supplied power cable. The power cable must
be protected against mechanical stress.
Magnetic and electrical fields can impair the function of the device. For this
reason, do not operate Cryo 7 in the vicinity of devices which generate strong
electromagnetic fields (X-ray or diathermy equipment, MRI machines). Please
keep a safe distance of several meters.
Cryo 7 is not suitable for use in areas with an explosive, flammable or
combustive atmosphere. The Cryo7 is intended to be used in closed and dust-
free treatment and examination rooms,
During use, Cryo 7 should be set up such that direct access to the central
mains power supply of the device is possible so that it can be disconnected
from the mains at any time.