Ziehm Imaging, Inc.
Vision Installation Manual
MAN 06-0015, Rev. A
Page 35
M a i n t e n a n c e R e p o r t : S m a l l e s t F i e l d S i z e
Fig. 24
Measuring the Smallest Field Size
7. Enter the Integrated Service Functions mode, select Program 1, and restore the
value to the one recorded earlier. Save these settings and exit the Integrated Ser-
vice Functions mode.
Smallest field size: Film (A) ________ cm (B) __________ cm
a
a. Reject if smallest field size is greater than 5 cm x 5 cm.