Ziehm Imaging, Inc.
Vision Installation Manual
MAN 06-0015, Rev. A
Page 42
M a i n t e n a n c e R e p o r t :
Cleaning,
Disinfecting, and
Sterilization
Turn the X-ray system off and disconnect it from the wall outlet power connection,
before cleaning or disinfecting it.
WARNING: Keep moisture out of the X-ray system!
DO NOT spray the system
with water or any kind of liquid ! DO NOT use a cloth soaked in any kind of liquid to
clean the equipment.
Cleaning
You may clean the X-ray system with water, standard household cleaners, and a damp
cloth. Do not use any scouring cleaners, organic thinners, or cleaners that contain thin-
ners (i.e. alcohol, pure benzene, stain remover).
To clean the display screens, only use pure alcohol, or a mixture of 1/3 alcohol to 2/3
distilled water.
Immediately after you clean the X-ray system and/or display screens, rub them dry
with a soft cotton cloth.
Disinfection
To disinfect the system, we recommend using the following disinfectants:
Incidin (3% in water)
Ultrasol F (5:1 in water)
We guarantee that Incidin and Ultrasol, used as described above, will not degrade the
X-ray system’s painted surfaces. We do not guarantee the use of other disinfectants
and/or substances on painted surfaces.
Sterilization
The removable cassette holder and cotton covers can be sterilized. Appropriate steril-
ization methods must be used to ensure the sterility of cotton covers. Ziehm does not
supply pre-sterilized cotton covers. The sterilization facility operator is responsible for
ensuring proper sterilizations methods are used.