Data input fields
•
Start the service software (see "
Opening
the Test Tools program"
).
•
Select the
data
tab.
Driver information field
type HW
instrument type selection.
DateEE
EEPOM programming date.
Serial Nr
.
entry of the serial number.
Date VCGIF
Grabber driver date.
Enter defaults
Enter the default values (factory settings) for
temperature levels and sensor.
After changing the entries,
click on the
save data
button.
Save data
The data is saved in the EEPROM of the VISUCAM
lite
control system.
Temperature levels field
These preset values must not be changed.
high temperature
⇒
fixed value 085 – corresponds with 50°C
critical temperature
⇒
fixed value 105 – corresponds with 60°C
•
The halogen lamp will be switched off.
•
An error message will be displayed.
Sensor field
This field shows the camera positions currently stored.
HP-Pos0
B/W camera sensor position in steps (steps between light barrier and B/W camera
sensor)
P0-P1
color camera sensor position in steps (steps between B/W camera sensor and
color camera sensor).