Service Instructions VISUCAM
lite
Issue: 22.09.2003
SM-30-4044-A1-en
Replaces issue:
Page
31
of 41
Opening the Windows Explorer program
⇒
⇒
⇒
⇒
VISUCAM
lite
is on and in the patient or image
capture mode.
•
Click the
?
in the program bar.
⇒
⇒
⇒
⇒
The window with Help Topics and Info opens .
•
Click
Info
⇒
⇒
⇒
⇒
The
About
windows opens.
This window contains important data on
VISUCAM
lite
including the serial number.
•
Press the
??? +?
key combination.
Note:
This key combination is only available to
service staff and is provided during service
training.
⇒
⇒
⇒
⇒
The Windows Explorer opens,
providing access to the system settings.
Closing an active window
Use the
ALT + F4
key combination to close an active window.
Opening watch files (Logserver)
Use the
ALT + w
key combination to open newest watch file (watch
x
.lgg) of the logserver.
Two more watch files are in the folder
C:\VISUCAM
\Watch(1-3).lgg. To open watchx.lgg is not
possible. X can be no. 1-3. The oldest watchx.lgg will be deleted after start, opened new by
logserver and get the newest data if VISUCAM
lite
program is active.
Enter BIOS
To enter BIOS, press
F2
several times during booting the instrument.
Password for BIOS
The Password is
????? (only available to service staff!)
Installing, uninstalling CD-writer
Install / uninstall new drivers for CD-writer (Plexwriter)
Uninstall: 1. IN CD program
Install:
1. Driver for the CD-writer
2. Driver for the CD-writer
2. IN CD program
Note:
If it is not possible to read any data from a CD-RW, you have to install the
UDF- Reader program on your computer to read files from CD-RW.
Drivers are local at c:\ installs