Service Instructions VISUCAM
lite
Issue: 22.09.2003
SM-30-4044-A1-en
Replaces issue:
Page
9
of 41
Help for troubleshooting
Questions about mechanical problems
•
Can the camera head be swiveled all the way to the right and left?
•
Can the VISUCAM
lite
fundus camera be easily moved and locked on the base within the
entire working range?
•
Is the ophthalmoscope lens clean?
Questions about electrical or system problems
•
Have all connections been made and all plugs been properly inserted?
•
Does the computer start up after the power switch has been switched on?
•
Can the operating software be started?
•
Can the brightness of the halogen lamp be adjusted?
•
Does the drive for the exciter filter initialize?
•
Does the drive for the barrier filter initialize?
•
Does the illumination of the finder camera function properly for white light and I R light?
•
Does the finder camera function properly and is the image provided by it displayed on the
monitor?
•
Can the B/W camera sensor (1/2”) be selected for observation?
•
Can the color camera sensor (1/2”) be selected for observation?
•
Does the focusing system function properly?
•
Does the B/W camera sensor function properly and can images be captured?
•
Does the color camera sensor function properly and can images be captured for 45°/35°?
•
Can you print with the printer (option) connected?
•
Have the screen settings (contrast, brightness) been adjusted as required?
•
Have the screen values (1024x768, 24bit, true color) been adjusted?
•
Does external archiving function (optional network or USB Flash Drive) properly?