Service Instructions VISUCAM
lite
Issue: 22.09.2003
SM-30-4044-A1-en
Replaces issue:
Page
10
of 41
VISUCAM
lite
- Component Service
Note:
Easy-to-exchange components are not described in detail.
Replacing the blinking diode of the fixation
light
•
Screw out the blinking diode (1) from its
socket(2).
•
Screw the new blinking diode into the socket.
•
Check whether it is properly functioning.
Opening and closing the camera head
•
Disconnect the VISUCM C from line
power.
•
Remove the lamp housing cover.
(Snap lock)
•
Remove Focus Drive left side and right
side.
•
Loosen 4 x hex socket head screws
(2.5 mm) in the left-hand cover.
•
Remove the left-hand cover.
•
Loosen 4 x hex socket head screws
(2.5 mm) in the right-hand cover.
•
Remove the right-hand cover.
Note:
When removing and attaching the
covers, make sure that the finder
camera cover plate is in the grooves
and not damaged and take care not
to squeeze any cables.
•
Assemble in reverse order.
Replacing the halogen bulb
Caution:
The halogen bulb may be hot. Risk of burns!
Allow the halogen bulb to cool down.
•
Open the camera head
(see: Opening and closing the camera head).
•
Remove the halogen bulb including socket
from the lamp mount.
Take care not to bend the mount!
•
Remove the defective halogen bulb from the
socket.
•
Insert the new halogen bulb into the socket.
•
Insert the halogen bulb including socket into
the mount.
•
Close the camera head
(see: Opening and closing the camera head).
•
Switch on VISUCAM
lite
and check its functions.
Note:
The lamp need not be readjusted
after bulb replacement.
1 Blink diode
2 Socket of Blink diode
Summary of Contents for VISUCAM lite