Service Instructions VISUCAM
lite
Issue: 22.09.2003
SM-30-4044-A1-en
Replaces issue:
Page
25
of 41
Exchanging Motherboard MEDP3
Note:
To be able to exchange motherboard MEDP3, you first have to remove instrument
interface GIF.
Motherboard connectors
Note:
Do not fix the motherboard
here with a screw!
Exchange hard disk
Note:
If possible, before exchanging the hard disk a backup of the patient data, network and
printer settings should be done.
•
Exchange the hard disk (see:
components of the control base
).
•
Restore the operating system and the data base using the recovery CD-ROM.
(SM-30-4045-A2).
•
Reinstall printer driver and run it (optional).
•
Check the position of the finder camera window (see:
checking and adjusting the image
position of the finder camera
).
•
Restore saved patient data.
•
Reinstall the network settings (optional).
Exchange GIF instrument interface
For exchange see:
components of the control base.
Connections of the GIF instrument interface
X10 Motherboard
X9 Motherboard
X21
X22
X3
X5
X6
X7
Sensor
left/right movement
Power
supply
Fixation lamp, head rest
Fan base
Camera head
X1
X2
X4
VGA Display
Con. panel
Hard disk
CD-ROM / burner
Power
supply
User interface
X21
User interface
X22
Battery
X19
X17
X9
X10
X6
PS2 Mouse key
board Con. panel
X5