Instrument Description
000000-1253-546 VISUCAM C 31.07.2003
14
Power isolation transformer for external devices
Caution
Make sure you always connect all external devices, printer and monitor,
to the power isolation transformer.
No other components than those system components described may be
connected to the power isolation transformer or instrument table.
Non-compliance presents a violation of the regulations for the use of
medical devices as per DIN EN 60601-1-1.
If you use the IT 3L or IT 2S Instrument Table from Carl Zeiss for the
VISUCAM C, the power isolation transformer can be mounted to the
underside of the tabletop. Otherwise, it must be set up safely, but not
on the floor.
1
2
1
Power cable connector with
fuses
2
Power switch
Fig. 5
Power isolation
transformer, front view
2
1
2
2
1
Appliance connector
2
Power junction connector
Fig. 6
Power isolation
transformer, rear view