66
Maintenance / Further information
G-30-1439-en
OPMI® VISU 200 BrightFlex™ Surgical Microscope
Issue 3.0
Printed on 21. 11. 2002
Sterilization
The asepsis sets available from Carl Zeiss contain rubber caps, sleeves
and handgrips which can be sterilized in autoclaves. We recommend the
following program for sterilization:
Sterile single-use drapes are available to cover the instrument.
Note:
When draping the system, make sure there is enough slack in the drapes
to allow for movement of the microscope carrier and surgical microscope.
It is especially important that the drapes are completely loose around the
handgrips. The surgeon must be able to operate the keys through the
drape.
Sterilization temperature:
134° C
Sterilization time:
10 minutes