52
Operation
G-30-1439-en
OPMI® VISU 200 BrightFlex™ Surgical Microscope
Issue 3.0
Printed on 21. 11. 2002
Setting the microscope tilt to angles greater than 15°
•
Using locking screw (2), secure the assistant's microscope in the
working position set (on the left or right of the main microscope).
•
Check that the assistant's microscope is firmly seated.
•
Remove the binocular tube and all accessories mounted on the assist-
ant's microscope.
•
Remove all accessories mounted on the objective lens (e.g. 0° or 8°
assistant's microscope).
•
Turn knob (1) until the surgical microscope is in the viewing position
required.
After the viewing angle has been set, the surgical microscope remains in
this position. The gear drive is self-locking.
2